Maintenance
61
Operating Instructions HYGROPHIL® H 4230-12 BA 060207
6
Maintenance
6.1
General information
HYGROPHIL
®
4230 is as good as maintenance-free. The only maintenance
necessary during operation is to refill the water tank and add a surfactant.
Depending on the conditions of use, it may be necessary for the measuring
chamber and any filters installed to be cleaned periodically.
The water separator on the external reducing valve should be checked oc-
casionally. If necessary, this must be emptied and the filter must be cleaned.
In rare cases when particular types of dirt occur, it is also necessary for the
gas sampling hose to be inspected and cleaned on a regular basis.
Particular attention must be paid to making sure that the entire air supply
system is completely leak-proof. There is always negative pressure in the
measuring chamber. Leaks cause leakage air to be inducted.
Even small quantities of leakage air seriously falsify the measurement result.
You must therefore make sure that the air supply system is completely leak-
proof.
Check hoses, lines and connections periodically for leakages, wear and tear
or damage. Repair any parts that need repairing.
The rest of this chapter provides maintenance instructions for the following
components of the equipment:
Analogue outputs
Measuring chamber
Water pump
Gas sampling hose
Reducing valve
Summary of Contents for HYGROPHIL H 4230-12 Ex Zone II
Page 2: ......
Page 30: ...System description 24 Operating Instructions HYGROPHIL H 4230 12 BA 060207...
Page 34: ...Safety precautions 28 Operating Instructions HYGROPHIL H 4230 12 BA 060207...
Page 66: ...Programming 60 Operating Instructions HYGROPHIL H 4230 12 BA 060207...
Page 112: ...Operating Instructions HYGROPHIL H 4230 12 BA 060207...