3.3
Connection instructions
3.3.1
Assignment of terminals
3.3.2
Mains connection
Wiring must be carried out exclusively by specialists who are familiar
with the valid local regulations for setting up power installation rated for
up to 1000 volts.
Check whether the voltage of the mains corresponds to that on the rat-
ing plates of the equipment and gas sampling hose.
All cables that are used must be selected in accordance with DIN/VDE
or the relevant national regulations.
Install an all-pole switch near the equipment to switch the equipment on
and off.
Run the mains cable through the (anti-kink) cable gland on the right.
Connect the mains voltage cable to terminals 27 (L), 28 (N) and 29 (PE).
Do not disconnect the yellow/green protective earth conductors or those
labelled „PE“ at any point.
Wire all power consumers within a complex system, including the protec-
tive earth conductors, in a star connection. Avoid sequential wiring.
Under no circumstances use the unit’s supply terminals as a terminal for
other loads.
In the case of interference-prone power supplies, it is a good idea to
connect an isolating transformer on line side, and also a mains filter if
required.
If there are several units in one system, each must be protected sepa-
rately against short circuits and equipped with a disconnector.
When the work is finished, tighten the threaded joints on the cable
glands. This ensures that the degree of protection is observed and pro-
vides strain relief.
Check that the protective earth conductor is connected correctly.
OUT 1 OUT 2
External
water
detector
Warning
relay
T1
external
Heating
Mains
+
IN
–
L N PE
+ –
+ –
Error
relay
L N PE
PROFIBUS
+5V A B 0V
Summary of Contents for HYGROPHIL H 4230-12 Ex Zone II
Page 2: ......
Page 30: ...System description 24 Operating Instructions HYGROPHIL H 4230 12 BA 060207...
Page 34: ...Safety precautions 28 Operating Instructions HYGROPHIL H 4230 12 BA 060207...
Page 66: ...Programming 60 Operating Instructions HYGROPHIL H 4230 12 BA 060207...
Page 112: ...Operating Instructions HYGROPHIL H 4230 12 BA 060207...