System description
13
Operating Instructions HYGROPHIL® H 4230-12 BA 060207
1.4
Technical data
Humidity measurement
Measurement principle
Psychrometric gas humidity measurement in line with the impact jet method
Transducer
PT 100/ 4-conductor in accordance with DIN IEC 751
Computational accuracy
0.01%
Computing time
Approx. 2s
Settling time
t
90
= 90s (for sudden change in SH from 10 to 190 g/kg)
Air/gas flow rate
Adjustable, max. 17,5 Nl/min
Water intake
Max. 25 ml/h (hose pump)
Water reserve
2l (enough for approx. 3 days)
Compressed air intake
2...5 bar (max. air consumption 2000 Nl/h) -when operating without over-pressure-
Measured variable inputs
Measured variable
Measurement range
Resolution
Accuracy
Type
Dry temperature TT
0...140 °C
0.1 °C
0.5% of the
measurement
range
Primary
Humid temperature HT
0...140 °C
Temperature T1
external
(optional)
0...200 °C
Absolute pressure SP
500...1500 hPa
1 hPa
1%
Dew-point temperature DT
20...100 °C
0.1 °C
Calculated
Volumetric content H
2
O Vol%
2...100 %
0.1 %
Absolute humidity MH
15...1000 g/kg
1 g/kg
Specific humidity SH
15...1000 g/kg
1 g/kg
Enthalpy h
35...1000 kJ/kg
1 kJ/kg
Current vapour pressure VP
10...1000 hPa
1 hPa
Saturation deficit DVP
0...1000 hPa
1 hPa
Outputs
Signal output
Analogue output
2 electrically isolated output channels, can be assigned to each of the measurement
ranges, spread, error behaviour programmable
Output signal
0…20 mA or 4…20 mA (programmable), linear
Permissible load
500
Accuracy
0.2% of the associated measured value
Inputs
External water detector
24 V d.c., NPN
T1
external
PT 100/ 4-conductor in accordance with DIN IEC 751
Data interface
Field bus interface
PROFIBUS DP
Electrical data
Auxiliary power
Measuring unit: 90…264 V a.c., 47...63 Hz, approx. 30 VA
Gas sampling hose: 230 V or 115 V a.c.; approx. 100 VA/m
Relay
Warning relay
Display of warnings Load: 1 A/24 V d.c., at least 10 mA
ERROR relay
Display of failures Load: 1 A/24 V d.c., at least 10 mA
Ambient conditions
Permitted working temperature +5...+50 °C
Permitted storage temperature -20...+70 °C (without water)
Climate category
KWF in accordance with DIN 40040
Reference conditions
23 °C
2 °C / 230V ± 2%
Summary of Contents for HYGROPHIL H 4230-12 Ex Zone II
Page 2: ......
Page 30: ...System description 24 Operating Instructions HYGROPHIL H 4230 12 BA 060207...
Page 34: ...Safety precautions 28 Operating Instructions HYGROPHIL H 4230 12 BA 060207...
Page 66: ...Programming 60 Operating Instructions HYGROPHIL H 4230 12 BA 060207...
Page 112: ...Operating Instructions HYGROPHIL H 4230 12 BA 060207...