PROFIBUS
100
Operating Instructions HYGROPHIL® H 4230-12 BA 060207
DaAdr $48: T2 correct config
Format: Char
Before you can write this variable you must first change to standby mode.
0x00 = No correction of HT based on the T2 measurement
0x80 = HT is corrected based on T2 (internal temperature) (default)
DaAdr $4c: Timeout activating
Format: Char
Before you can write this variable you must first change to standby mode.
7
6
5
4
3
2
1
0
ON
free
free
free
free
free
free
free
ON:
1 = timeout monitoring active
0 = no timeout
8.3.6
2-channel parameters
DaAdr (hex) Identifier
Memory-Type
Read-out Format
$50
PropValveControlStatus
RAM read/write
hex
Record
$54
PumpControlStatus
RAM read/write
hex
Char
$55
PumpConfig
EEPROM read/write
hex
Char
$57
EquipmentControlStatus
RAM read/write
hex
Char
$58
EquipmentConfig
EEPROM read/write
hex
Char
$53
Limit Task config.
EEPROM read/write
hex
Char [2]
$5a
Limit LOW
EEPROM read/write
dec
Float
$5b
Limit HIGH
EEPROM read/write
dec
Float
$5c
DP Parameter
EEPROM read/write
hex/dec
Record
SWNo $50: PropValveControlStatus
Format: Record
Statuschar: Char;
ControlStatusChar: Char;
ON time; Integer;
OFF time; Integer;
End;
StatusChar:
7
6
5
4
3
2
1
0
ON
OpenError CloseError Free
Free
Free
Free
Free
ON:
1 = Controller is switched on and working
0 = Controller off
OpenError: 1 = Flow rate too low, valve cannot be opened any further
0 = Valve in working range < maximum
CloseError 1 = Flow rate too high, valve cannot be closed any further
0 = Valve in working range > minimum
Summary of Contents for HYGROPHIL H 4230-12 Ex Zone II
Page 2: ......
Page 30: ...System description 24 Operating Instructions HYGROPHIL H 4230 12 BA 060207...
Page 34: ...Safety precautions 28 Operating Instructions HYGROPHIL H 4230 12 BA 060207...
Page 66: ...Programming 60 Operating Instructions HYGROPHIL H 4230 12 BA 060207...
Page 112: ...Operating Instructions HYGROPHIL H 4230 12 BA 060207...