
7.4
Strain Gages
Due to variations in materials, manufacturing and external forces, the strain gage values may
change. These changes will affect the zero force reading for each beam differently. To maintain
consistent results, the zero force reading needs to remain constant. Each strain gage is equipped
with a balancing potentiometer. Adjusting the balancing potentiometer will change the strain gage
output for that finger. Adjust the BH8-262 balancing potentiometer until the no-load value is
between 100 and 140. Use the following steps to zero the strain gages:
1.
Initialize the BarrettHand™.
2.
Terminate the spread motor so it can be moved around the palm. (Issue the "T" command)
3.
Remove the Shroud Cover screws shown in Figure 23. Some models of the BarrettHand™
will have four Shroud Cover screws. Remove the Shroud from the finger link.
Figure 23 - Shroud Removal
4.
BH8-262 users should run the program
Monitor Strain
. This program will continuously
sample the strain gage values and print them to the screen. Another option is to run the
BarrettHand Control GUI and poll for strain values.
5.
Adjust the balancing potentiometer using a small flat head screwdriver until the desired value
is reached. The balancing potentiometer requires very small adjustments, due to its
sensitivity. Apply as little pressure as possible on the balancing potentiometer during
adjustment. See Figure 24.
Page 56 of 89
Shroud Cover
Screws
Shroud Cover