Summary of Contents for MaxQ 2000 Series

Page 1: ...MaxQ 2000 Open Air Platform Shakers MODEL NO SHKA2000 SHKE2000 SHKA2000 1CE SHKE2000 1CE 057 278 00 1 17 05 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 2: ...p Installation 13 Test Tube Rack Installation 13 Operation 14 SHKA2000 SHKA2000 1CE 14 Power Switch 14 Speed Control and Display 14 Time r 14 SHKE2000 SHKE2000 1CE 15 Turning Shaker On 15 Setting Shak...

Page 3: ...ignite such materials Not rated for use in hazardous atmospheres 2 Use appropriate hand and eye protection when handling hazardous chemicals 3 Refer servicing to qualified personnel 4 Do not modify co...

Page 4: ...xing etc The vessels should be securely clamped before turning on shaker and contain a stopper whenever possible to prevent hazardous sub stances from being thrown out during shaking action Shaking ac...

Page 5: ...terface None None Monitor speed and time with a computer Monitor speed and time with a computer Display Analog tachometer displays speed in rpm Analog tachometer displays speed in rpm 2 individual LED...

Page 6: ...s Beeps twice when time has expired Shaking motion stops Unbalanced Load None None If the unit is running in an unbalanced condition an alarm will sound and the shaker will stop until the end user cor...

Page 7: ...eypad in one rpm increments Flashing display indi cates power interruption Pressing any key will clear display Non volatile memory maintains speed and time set points in the event of a power interrupt...

Page 8: ...ions and flicker EN 61326 1 Electrical equipment for measurement control and laboratory use Part I General Requirements Safety EN 61010 1 Safety requirements for electrical equipment for measurement c...

Page 9: ...d 3 Speed Tachometer Analog display of platform rotation speed 4 Time r Allows user to choose either continuous or timed operation Caution It s not recommended to operate shaker in a CO2 enriched atmo...

Page 10: ...digit LED indicates time remaining or elapsed time 11 Up Arrow Key Increases shaking time 12 Down Arrow Key Decreases shaking time 13 Elapsed Timer Light Indicates accumulated time 14 Hours Timer indi...

Page 11: ...below when unpacking to verify that the complete unit has been received Do not discard packing materials until all is accounted for The following items are included in the shipment Model No SHKA2000...

Page 12: ...and or neutralize the spilled liq uids before continuing Leave shaker disconnected when not in use Platform Installation 1 Select the appropriate platform for the vessels to be shaken A wide variety o...

Page 13: ...provided with the clamp must be properly attached to the platform 2 Make sure all vessels are securely clamped before turning on unit Wherever possible vessels should contain a stopper to prevent haza...

Page 14: ...a maximum of 400 rpm Time r 1 From the 12 o clock off position rotate timer knob counterclockwise to the ON position to ini tiate continuous operation 2 For timed operation rotate timer knob clockwise...

Page 15: ...ress STOP membrane switch to end shaking SET RPM light will illuminate Calibrating Shaking Speed 1 Choose a speed for which calibration is desired by using the shaker s UP or DOWN arrow keys 2 Measure...

Page 16: ...ne switch for desired timing mode 3 Hold down appropriate arrow membrane switch in the TIME module of the control panel up or down until desired timing cycle is set from 0 1 hour up to 999 hours or fr...

Page 17: ...timed shaking Timer will begin to count up and will display accumulated time in display window Platform rotation will cease and TIME display will flash when ELAPSED time reaches 999 minutes or 999 hou...

Page 18: ...None Stop bits 1 Flow control None After verifying the above settings click OK 6 In the main dialog box click on File Save 7 Exit the program by clicking on File Exit Yes 8 Verify the program was save...

Page 19: ...atically set by the location of the motor and is not adjustable 4 To align a pulley loosen the motor pulley set screw slide the pulley up or down into alignment then securely tighten the motor pulley...

Page 20: ...i n u s e h c t a m t a h t y l p p u s r e w o p e h t e t a c o L s t n e m e r i u q e r e s u f e c a l p e R l e n a p l o r t n o c n o h c t i w s e n a r b m e m y n a s s e r P r e m i t t e...

Page 21: ...HKA2000 1CE 562 184 10 Thumb Screw Knob 790 078 00 Mounting Feet 790 316 11 Mounting Plate Mat 019 455 00 Shaker Mechanism 370 272 01 Motor Brush Spring Replacement SHKA2000 SHKA2000 1CE 019 533 02 Ma...

Page 22: ...tained promptly from the factory When service or replacement parts are needed we ask that you check first with your dealer If the dealer cannot handle your request then contact our Customer Service De...

Page 23: ...23 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 24: ...d without further force or effect if there is any i repair made to the product by a party other than Barnstead or its duly authorized service representa tive ii misuse including use inconsistent with...