Carbon Filter Installation
Chlorine can also damage your membranes, resulting in
premature membrane failure. Therefore, your DIamond TII
uses an extruded carbon filter to remove chlorine from your
feed water.
Install the carbon filter as follows:
1.
Remove the carbon filter (Part Number D502115)
from its packaging and wet the o-rings with water.
2.
Insert the carbon filter straight into the right posi-
tion upper and lower quick disconnect and push
until firmly connected. For easier installation, con-
nect bottom first.
RO Membrane Installation (See Fig. 7 - 10)
12 LITER PER HOUR UNITS (1 MEMBRANE)
1. Remove the RO membrane from its packaging.
Note the location of the FEED, PRODUCT, and
REJECT connections on the membrane. The cor-
rect orientation for the installed membrane will be
vertical, with the Feed connection on the bottom.
See Figure 8.
2.
Open the left side door on the DIamond TII.
3. Referencing Figure 7, locate the three unconnect-
ed tubes (two with elbows, one without) labeled
FEED (A), PRODUCT (B) and REJECT (C). Refer
to “Tubing Installation: Figure 2” and mark FEED
(A) tubing 3/4” (1.9 cm) from end.
4. Make FEED (A) connection by first wetting the
tube with water and pushing it firmly into the mem-
brane connector. See Figure 7.
5. Using the Velcro
®
strap secure the membrane into
the right most position against the bracket.
6. Make the PRODUCT (B) and REJECT (C) connec-
tions by first wetting the tube stem of the elbow
with water and pushing it firmly into the membrane
connector. See Figure 7.
I
NSTALLATION
Figure 7: 12 lph RO Membrane Installation
UV Lamp
Velcro
®
is a registered trademark of
Velcro Industries B.V. Ltd Liab Co. Netherlands
Castorweg 22-24 Curacao Netherlands Antilles
18