11
Locating Unit
Locate storage reservoir approximately 5 ft. (1.52 m) from
the DIamond TII for electrical connections. See DIamond
TII storage reservoir manual or customer-supplied reser-
voir manual for mounting instructions.
The outlet of
DIamond TII storage reservoir must be above or at
the same level as the inlet of the DIamond TII.
Bench Mounting
Place the DIamond TII on a bench top that is accessible
to electricity and an atmospherically vented drain.
Wall Mounting
Install the DIamond TII on a wall in a convenient location
that is accessible to an atmospherically vented drain and
electricity.
1.
Locate the wall bracket packed separately from
the unit.
2.
Using the wall bracket as a template, locate and
drill the mounting holes in the wall. A minimum
of four (customer-supplied) fasteners will be
required — two on the top and two on the bot-
tom.
3.
Attach the wall bracket to the wall using cus-
tomer-supplied fasteners.
4.
Remove the locking screws on each side of the
wall bracket.
5.
Pull the two locking slides on each side of the
wall bracket out as far as they will go.
6.
Hang the unit on the wall bracket by sliding the
mounting pins into the wall bracket slots.
7.
Push the locking slides on each side of the wall
bracket in as far as they will go.
8.
Replace the locking screws.
Installation
Note
Please refer to the “General
Specifications” section for clearance
requirements.
Warning
Do not locate the DIamond TII directly
over equipment that requires electrical
service. Routine maintenance of this
unit may involve water spillage and
subsequent electrical shock hazard if
improperly located.
NOTE
The outlet of DIamond TII storage
reservoir must be above or at the
same level as the inlet of the
DIamond TII.
********************
************************
Caution
Wall composition, condition and con-
struction, as well as fastener type,
must be considered when mounting
this unit. The mounting surface and
fasteners selected must be capable of
supporting a minimum of 150 lbs.(68
kg). Inadequate support and/or fasten-
ers may result in damage to mounting
surface and/or equipment. If you are
unsure of mounting surface composi-
tion, condition and construction or cor-
rect fasteners, consult your building
maintenance group or contractor.