Operation
Conditions
Program
FFP
Dimensions
Filling vol-
ume pe r FFP
Loaded with
1 FFP
Loaded with
2 FFP´s
Loaded with
3 FFP´s
Loaded with
4 FFP´s
PLASMA
(approx. 170 x
110 x 20 mm)
no additional
packaging
250ml
Heating duration in minutes to reach 5°C
System te mperature
Number of FFP´s (see Table 1 for layout)
1
2
3
4
45°C
11
11
12
12
37°C
14
14
15
15
Heating duration in minutes to reach 33°C
System te mperature
Number of FFP´s (see Table 1 for layout)
1
2
3
4
45°C
14
14
16
20
37°C
22
22
24
30
4.4.8
Presetting the heating duration and starting the "BLOOD" program
WARNING
Plasma, blood and HPC may only be heated for a limited period of time!
Do not use the device to keep these products warm indefinitely!
WARNING
Blood/whole blood may not be heated up if the current operating temperature of
the device is over 37 °C. The device will display a warning message in this case.
The process for presetting the heating duration and starting the "PLASMA" program is
described in the Chapters Selecting and starting the “PLASMA” program and Messages
at the start of a program; to preset the heating duration and start the "BLOOD" program,
please proceed in the same way but select "BLOOD" instead of "PLASMA" as the pro-
gram.
4.4.9
Presetting the heating duration and starting the "HPC" program
WARNING