Safety
advice
3.3
Safety advice on environmental influences
WARNING
The influence of strong electromagnetic fields (e.g. through the
use of HF therapy or surgical devices) can lead to malfunctions
in the Barkey plasmatherm. If interference of this type occurs,
increase the distance between the Barkey plasmatherm and the
device causing the interference, or operate the devices at dif-
ferent times. The Barkey plasmatherm works perfectly within
the limit values set in the EN 60601-1-2 standard. The device
can be influenced outside the limit values set by EN 60601-1-
2.
Portable and mobile HF communication equipment such as
mobile phones can affect the device.
Do not use the device in the immediate vicinity of
flammable materials (e.g. gases, liquids),
flammable mixtures of anaesthetic substances with air,
flammable mixtures of anaesthetic substances with oxygen
or nitrous oxide
whose flashpoint is below 50°C. It is imperative that the device
is not used in areas in which alcohol disinfectants and anaes-
thetising substances are being used simultaneously.
The device may not be set up or operated in the immediate
vicinity of devices with a high heat output.
The device must be positioned so as to ensure an unrestricted
flow of air around its base.