Safety
advice
Gui delines and manufacturers declar ati on - Electr omagnetic i mmunity
The Ba rkey p lasmatherm is intended for operation in one of the e lectro magnetic enviro n-
ments listed below. The customer or user of the Barkey plas matherm must ensure that it is
operated in one of these environments.
Immunity test
IEC 60601
test le vel
Compliance le vel
Electromagne tic envir onme nt -
guideline
Conducted HF
interference ac-
cording to IEC
61000-4-6
Radiated HF in-
terference accord-
ing to IEC 61000-
4-3
3 V
e ff
150 kHz to
80 MHz
3 V/ m
80 MHz to
2.5 GHz
3 V
e ff
150 kHz to 80
Mhz
3 V/ m
80 MHz to 2.5
GHz
Portable and mobile rad io sets should
not be used within a distance fro m the
Barkey plas matherm, including cords,
that is less than the recommended
safety distance as calculated by the
relevant equation for the transmit
frequency.
Recomme nde d safety distance
GHz
2.5
to
MHz
800
for
33
.
2
MHz
800
to
MHz
80
for
17
.
1
MHz
80
to
kHz
150
for
17
.
1
P
d
P
d
P
d
where
P
is the nominal powe r of the
transmitter in Watts (W) as stated by
the transmitter manufacturer, and
d
is
the recommended safety distance in
metres (m).
The fie ld strength of stationary
transmitters should always be less
than the compliance leve l
b
at all
frequencies in accordance with an on-
site investigation
a
.
Interference is possible near devices
that display the following symbol.
NOTE 1 At 80 M Hz and 800 MHz the higher frequency range applies.
NOTE 2 These guidelines may not apply in all cases. The propagation of electromagnetic
variables is affected by absorption and reflection of buildings, objects and people.
a
The fie ld strength of static transmitters, such as base stations of radio telephones and
mobile land radios, a mateur radios, AM and FM radio and telev ision transmissions cannot be
accurately theoretically determined in advance. You should consider carrying out a site survey
to determine the e lectro magnetic environ ment with regard to static transmitters. If the meas-
ured fie ld strength at the location where the Ba rkey plasmatherm is being used exceeds the
above compliance levels, the Ba rkey p lasmatherm should be monitored to ensure that the
device is functioning as intended. If unusual performance characteristics are observed, add i-
tional measures such as changing the align ment or location of the Ba rkey p lasmatherm may be
necessary.
b
The field strength should be less than 3 V/ m in the frequency range 150 kHz to 80 M Hz.