![Barkey 212.10043 Instructions For Use Manual Download Page 67](http://html1.mh-extra.com/html/barkey/212-10043/212-10043_instructions-for-use-manual_475147067.webp)
Specifications
11
Specifications
Manufacturer
Barkey GmbH & Co. KG
Gewerbestrasse 8
33818 Leopoldshoehe
Tel.: +49 (5202) 9801-0
Fax: +49 (5202) 9801-99
Type
Barkey plasmatherm
Order No.
212.10043
230 VAC / 50 - 60 Hz
212.10044
115 VAC / 60 Hz
Typical capacity
4 bags / 500 ml or
2 bags / 800 ml
Emergency: 8 bags / 250 ml
Power supply
230 VAC / 50 - 60 Hz or 115 VAC / 60 Hz
Typical power consumption < 0.25 W
Standby
< 35 Watt
Program pause
< 75 W
Program "PLASMA" 45°C
1600 Watt
max.
Sound pressure level
< 45 dB(A)
Current consumption
7 A at 230 VAC
max.
14 A at 115 VAC
max.
Fusing
8 A slow at 230 VAC
15 A slow at 115 VAC
Battery type
Lithium CR 1225, 3 V
Interfaces/devices
optional:
Barcode scanner
Log printer
Overtemperature protection software controlled:
se 1.0 °C
electronic:
48 ± 1.0 °C
Temperature setting
Variable from
+37 to +45 °C
Default setting
+37 °C
Display accuracy
0.1 °C
Operation
Storage/transport
Ambient temperature
+10 to +40 °C
-20 to +70 °C
Humidity
30 to 75 %
30 to 90 %
(non condensing)
Air pressure
700 to 1060 hPa
700 to 1060 hPa
Operating mode
Continuous