Error
messages
7.1
Moisture sensors
If any liquid has leaked inside the device heating chamber, the moisture sensors in the
base of the chamber trigger an intermittent acoustic signal and a warning message flashes
in the display.
►
Press the confirm button ("OK").
►
Switch off the device.
►
Clean, disinfect and dry the heating
cushion and the heating chamber.
►
Replace the filter paper (dry-paper).
Figure 39: Moisture sensor message
7.2
Overtemperature
The error message is initiated when the device detects an overtemperature condition.
►
Note the error message.
►
Make a note of the error number (xxx)
and serial number (SN).
►
Switch off the device and allow it to
cool down.
Figure 40: "Overtemperature!" error message
An overtemperature condition can occur for a variety of reasons. Further diagnosis is pos-
sible by reference to the error number (see also the Chapter Error numbers). In order to be
certain that there is an error in the device, you must switch it on again after allowing it to
cool down.
If the error message reappears, please contact your medical technical service or the ser-
vice department of Barkey GmbH & Co. KG stating the error number and the serial
number.
! E r r o r!
O v e r te m p e ra t u r e !
E r r o r n u m b er : x x x
S N : 1 2 3 45 6 7
S w i t c h o f f t h e
d e v i c e
H e a t in g c us h i o n
o r c o n s er v e
h a s l e a ke d !
O K