Operation
4.4.3
Messages at the start of a program
Blood/whole blood should not be heated if the current operating temperature of the device
is over 37 °C, for example if you have previously thawed and heated plasma and you then
activate a program for heating blood/whole blood. You can nevertheless still start the
"BLOOD" program.
►
Acknowledge the display with the
left-hand selection button ("Yes") to
start the program despite the water
temperature being too high.
►
Acknowledge the display with the
right-hand selection button ("No") if
you do not wish to start the program.
Figure 15: Message "Water temperature too high!"
4.4.4
Messages while a program is running
If the heating chamber cover is opened while a program is in progress, the remaining time
is 'frozen', a message appears in the display and heat transfer fluid circulation is paused.
An intermittent acoustic signal sounds if the cover is not closed within 16s.
►
Close the heating chamber cover to
resume the program.
Figure 16: Message "Close the cover!"
NOTE
If paddle motion was active when the cover was opened, it will not resume until the hea t-
ing cushions have sufficiently filled again.
B L O OD
3 7 . 0 °C
W a t e r t e m pe r a t u re
to o h ig h !
= > 4 1 . 6° C
S t a r t t h e p r o g r am ?
Y e s N o
P L A S MA
C l o s e t h e co v e r !
R e m a i n in g t im e 1 0 :3 1
- S t o p +