![Barkey 212.10043 Instructions For Use Manual Download Page 39](http://html1.mh-extra.com/html/barkey/212-10043/212-10043_instructions-for-use-manual_475147039.webp)
Operation
►
Acknowledge the message with the
confirm button ("OK")
►
Top up with water as described in
Chapter 5.3.2
Figure 21: "Lack of water" message at the end of a programme
NOTE
This message is also output if the program is cancelled manually or as the result of an er-
ror.
NOTE
If this message is acknowledged 4 times without water being topped up, then an active
program will be immediately cancelled the next time low water is detected (see the Chap-
ter Messages while a program is running).
4.4.6
Presetting the heating duration of the "PLASMA" program
This chapter describes how to preset the heating duration, taking the "PLASMA" program
as an example; to preset the heating duration of the "BLOOD", "HPC" and "USER" pro-
grams, please proceed as described for the "PLASMA" program.
WARNING
Plasma, blood and HPC may only be heated for a limited time! You must not use
this device for keeping these products warm indefinitely!
►
In the main screen, press the confirm button ("Menu") to start the menu system.
L a c k o f wa t e r !
F I L L t a n k
O K
Summary of Contents for 212.10043
Page 2: ......
Page 31: ...Operation Figure 10 Optional loading positions for conserves...
Page 70: ......
Page 71: ......