6. Important information
Power connection – Equipment with internal power supply
•
This equipment must be earthed.
•
Power requirements: The equipment must be powered by the AC mains voltage.
•
The equipment should be installed near an easily accessible outlet.
•
The equipment is intended for continuous operation.
Power cords:
•
Utilize a UL-listed detachable power cord, 3-wire, type SJ or equivalent, 18 AWG min., rated 250 V
min., provided with a hospital-grade type plug 5-15P con
fi
guration for 120V application, or 6-15P for
240V application.
•
Do not overload wall outlets and extension cords as this may result in
fi
re or electric shock.
•
Mains lead protection (U.S.: Power cord): Power cords should be routed so that they are not likely to
be walked upon or pinched by items placed upon or against them, paying particular attention to cords
at plugs and receptacles.
•
Use a power cord that matches the voltage of the power outlet, which has been approved and complies
with the safety standard of your particular country.
Connections
Any external connection with other peripherals must follow the requirements of clause 16 of IEC60601-1
3rd ed. or Table BBB.201 of IEC 60601-1-1 for the medical electrical systems.
Connection of PEMS by network/data coupling to other equipment (MDSC-8358
version only)
•
Connection of the PEMS to a network/data coupling that includes other equipment could result in
previously unidenti
fi
ed risks to patients, operators or third parties.
•
The responsible organization should identify, analyze, evaluate and control these risks.
•
Subsequent changes to the network/data coupling could introduce new risks and require additional
analysis.
•
Changes to the network/data coupling include:
- changes in network/data coupling con
fi
guration;
- connection of additional items to the network/data coupling con
fi
guration;
- disconnecting items from the network/data coupling con
fi
guration;
- update of equipment connected to the network/data coupling con
fi
guration;
- upgrade of equipment connected to the network/data coupling con
fi
guration
Water and moisture
The equipment is IP20 compliant (IP21 for versions with protective glass).
Moisture condensation
Do not use the monitor under rapid temperature and humidity change conditions and avoid direct contact
with cold air from air-conditioning outlet.
Moisture may condense on the surface or inside the unit, which may cause a mist residue on the front
glass. Although this is not a malfunction of the product itself, it may cause damage to the monitor.
If condensation appears, unplug the monitor from the power source and allow enough time to restore to
its normal condition (i.e. until condensation disappears) before switching the monitor back on.
Ventilation
Do not cover or block any ventilation openings in the cover of the set. When installing the device in a
cupboard or another enclosed location, heed the necessary space between the set and the sides of the
cupboard.
26
(451920612552) K5902150 MDSC-8358 28/11/2017
Summary of Contents for MDSC-8358 RL
Page 4: ...Table of contents 2 451920612552 K5902150 MDSC 8358 28 11 2017 ...
Page 20: ...4 Daily operation 18 451920612552 K5902150 MDSC 8358 28 11 2017 ...
Page 26: ...5 Advanced operation 24 451920612552 K5902150 MDSC 8358 28 11 2017 ...
Page 42: ...6 Important information 40 451920612552 K5902150 MDSC 8358 28 11 2017 ...