3. Display installation
C
AUTION
:
The arm or stand is not provided. The product stability must be veri
fi
ed in
the speci
fi
c usage (according to IEC 60601-1 Clause 9.4.2.2)
To mount the display to an arm stand
1. Attach the mounting arm or the stand
fi
rmly
to the display using minimum 6 screws (M6) pointed out by
the red arrows in the image below. The screws can be inserted maximum 11 mm and minimum 8 mm.
Image 3-1
3.3
Video input connection
W
ARNING
:
When the display is assembled in the medical system, take care of the an-
chorage of all cables, to avoid unwanted detachment.
To connect the video inputs
1. Connect the available display(s) to the corresponding video input(s) using the appropriate video ca-
ble(s).
12
(451920612552) K5902150 MDSC-8358 28/11/2017