Mission critical applications
We strongly recommend there is a replacement monitor immediately available in mission critical applications.
Use of Electrical Surgical Knives
Provide as much distance as possible between the electrosurgical generator and other electronic equipment
(such as monitors). An activated electrosurgical generator may cause interference with them. The interference
can activate the OSD menu of the display and as such disrupt the functionality of the display.
Power connection
–
Equipment with internal power supply
•
This equipment must be earthed.
•
Power requirements: The equipment must be powered by the AC mains voltage.
•
The equipment should be installed near an easily accessible outlet.
•
The equipment is intended for continuous operation.
Transient over-voltage
To fully disengage the power to the device, please disconnect the power cord from the AC inlet.
Connections
•
Any external connection with other peripherals must follow the requirements of clause 16 of IEC60601-1
3rd. Ed. or Table BBB.201 of IEC 60601-1-1 for the medical electrical systems.
•
To maintain compliance with EMC Regulation, use only shielded interface cables for the connection to
peripheral devices.
Connection of PEMS by network/data coupling to other equipment (MDSC-8255 MNA
only)
•
Connection of the PEMS to a network/data coupling that includes other equipment could result in
previously unidentified risks to patients, operators or third parties.
•
The responsible organization should identify, analyze, evaluate and control these risks.
•
Subsequent changes to the network/data coupling could introduce new risks and require additional
analysis.
•
Changes to the network/data coupling include:
-
changes in network/data coupling configuration;
-
connection of additional items to the network/data coupling configuration;
-
disconnecting items from the network/data coupling configuration;
-
update of equipment connected to the network/data coupling configuration;
-
upgrade of equipment connected to the network/data coupling configuration
Power cords
•
Europe: H05VV-F or H05VVH2-F PVC cord with appropriate EU plug.
•
US and Canada:
“
hospital grade
”
cord-set has to be used, provided with instructions to indicate that
grounding reliability can be achieved only when the equipment is connected to an equivalent receptacle
marked hospital only or hospital grade. These instructions need to be marked either on the equipment or
on a tag on the power cord.
Grounding reliability
Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle.
Water and moisture
The equipment is IP20 compliant. The monitor front side is IPX5 compliant.
Moisture condensation
•
Do not use monitor under rapid temperature and humidity change condition or avoid cold air from air-
conditioning outlet directly.
Summary of Contents for MDSC-8255 12G
Page 6: ...R5913964 00 MDSC 8255 6 ...
Page 7: ...7 R5913964 00 MDSC 8255 Welcome 1 ...
Page 10: ...R5913964 00 MDSC 8255 10 Welcome ...
Page 11: ...11 R5913964 00 MDSC 8255 Parts controls and connectors 2 ...
Page 18: ...R5913964 00 MDSC 8255 18 Parts controls and connectors ...
Page 19: ...19 R5913964 00 MDSC 8255 Installation 3 ...
Page 26: ...R5913964 00 MDSC 8255 26 Installation ...
Page 27: ...27 R5913964 00 MDSC 8255 Daily operation 4 ...
Page 32: ...R5913964 00 MDSC 8255 32 Daily operation ...
Page 49: ...49 R5913964 00 MDSC 8255 Important information 6 ...
Page 70: ...R5913964 00 MDSC 8255 70 Important information ...
Page 71: ......