31
R5913216 /03
MDPC-8127
3.
Select
sRGB
and confirm.
4.13 Ambient Light Compensation (ALC)
About ALC
Ambient Light Compensation (ALC) can only be enabled on your display when the display function
is set to DICOM. Please refer to
, page 30
.
When ALC is enabled, the DICOM display function will be recalculated taking a preset ambient light correction
value into account. This value is determined by the selected reading room. Therefore, it is also important to
select a realistic reading room when enabling ALC. This can be done by following the instructions in
, page 31
To enable/disable ALC:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Ambient Light
menu.
3.
Enter the
Ambient Light Compensation
submenu.
4.
Select
Enabled/Disabled
as desired and confirm.
4.14 Reading rooms
About reading rooms
Reading rooms can only be selected on your display when the display function is set to DICOM.
Please refer to
, page 30
The American Association of Physicists in Medicine (AAPM) composed a list of pre-defined reading rooms.
Each of these reading rooms are defined by following parameters:
•
the maximum light allowed in this type of room
•
the preset ambient light correction value for this reading room
These parameters are stored in your display and determine the preset ambient light correction value to take
into account to recalculate the DICOM display function when Ambient Light Compensation (ALC) is enabled.
Please refer to
Ambient Light Compensation (ALC)
, page 31
to enable ALC.
The available reading rooms for your display are:
•
CR/DR/ MAMMO:
Corresponds to light conditions in diagnostic reading rooms for computed radiology,
digital radiology or mammography. This setting has the lowest maximum ambient light.
•
CT/MR/NM:
Corresponds to light conditions in diagnostic reading rooms for computed tomography,
magnetic resonance or nuclear medicine scans.
•
Staff Office:
Corresponds to light conditions in diagnostic reading rooms for digital pathology and office
rooms.
•
Clinical Viewing Room:
Corresponds to light conditions in diagnostic reading rooms for clinical viewing.
•
Emergency Room:
Corresponds to light conditions in emergency rooms.
•
Operating Room:
Corresponds to light conditions in operating rooms. This setting has the highest
maximum ambient light.
To select a reading room:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Ambient Light
menu.
3.
Enter the
Reading Room
submenu.
Summary of Contents for MDPC-8127
Page 5: ...5 R5913216 03 MDPC 8127 Welcome 1...
Page 12: ...R5913216 03 MDPC 8127 12 Welcome...
Page 20: ...R5913216 03 MDPC 8127 20 Installation and setup...
Page 21: ...21 R5913216 03 MDPC 8127 Daily operation 3...
Page 24: ...R5913216 03 MDPC 8127 24 Daily operation...
Page 37: ...37 R5913216 03 MDPC 8127 Cleaning your display 5...
Page 39: ...39 R5913216 03 MDPC 8127 Important information 6...
Page 51: ......
Page 52: ...www barco com R5913216 03 2021 09 10 Barco NV President Kennedypark 35 8500 Kortrijk Belgium...