R5913216 /03
MDPC-8127
30
2.
Navigate to the
Configuration > Calibration
menu.
3.
Enter the
Viewing Mode
submenu.
4.
Select
Diagnostic/Text
as desired and confirm.
4.11 Display functions
About display functions
Native, uncorrected panels will display all grayscale/color levels with luminance increments that are not
optimal for crucial diagnostic information. Studies have shown however, that in medical images certain
grayscale/color parts contain more diagnostic information then others. To respond to these conclusions,
display functions have been defined. These functions emphasize on these parts containing crucial diagnostic
information by correcting the native panel behavior.
The available display functions for your display are:
•
Native:
The native display panel behavior will not be corrected.
•
Dynamic Gamma 1.8
or
2.2
: These are gamma functions that are shifted to take into account the non-zero
luminance of an LCD panel when driven with a
“
black
”
signal. They are especially useful in CT applications
to improve the perception of low Hounsfield values.
•
DICOM:
DICOM (Digital Imaging and Communications in Medicine) is an international standard that was
developed to improve the quality and communication of digital images in radiology. In short, the DICOM
display function results in more visible grayscales in the images.
•
sRGB:
The sRGB color space combines a display function, color space and white point selection, and is
commonly used in digital pathology and is the default setting for the MDPC-8127.
Note:
This setting does
not fully set your display to the sRGB color space. Refer to
“
, page 30
to do this.
•
User:
This display function will be automatically selected when display functions are centrally defined by
QAWeb Enterprise.
•
Gamma 1.8
or
2.2
: Select one of these display functions in case the display is to replace a CRT display
with a gamma of 1.8 or 2.2 respectively.
•
Test:
For Barco service purposes only.
The settings of the display must be adapted to suit the requirements of the visualization software. In
case of doubt, please contact the vendor of the visualization software.
To select a display function:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration
menu.
3.
Enter the
Display Function
submenu.
4.
Select one of the available display functions and confirm.
4.12 sRGB
About sRGB
When sRGB is enabled, the display is set to an sRGB color space with the following settings:
•
Color preset:
6500K. See
, page 28
for more info.
•
Color space:
sRGB. See
, page 29
for more info.
•
Display function:
sRGB. See
, page 30
for more info.
To enable sRGB:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration
menu.
Summary of Contents for MDPC-8127
Page 5: ...5 R5913216 03 MDPC 8127 Welcome 1...
Page 12: ...R5913216 03 MDPC 8127 12 Welcome...
Page 20: ...R5913216 03 MDPC 8127 20 Installation and setup...
Page 21: ...21 R5913216 03 MDPC 8127 Daily operation 3...
Page 24: ...R5913216 03 MDPC 8127 24 Daily operation...
Page 37: ...37 R5913216 03 MDPC 8127 Cleaning your display 5...
Page 39: ...39 R5913216 03 MDPC 8127 Important information 6...
Page 51: ......
Page 52: ...www barco com R5913216 03 2021 09 10 Barco NV President Kennedypark 35 8500 Kortrijk Belgium...