5.1 Safety information
General recommendations
Read the safety and operating instructions before operating the device.
Retain safety and operating instructions for future reference.
Adhere to all warnings on the device and in the operating instructions manual.
Follow all instructions for operation and use.
Electrical Shock or Fire Hazard
To prevent electric shock or fire hazard, do not remove cover.
No serviceable parts inside. Refer servicing to qualified personnel.
Do not expose this apparatus to rain or moisture.
Modifications to the unit
Do not modify this equipment without authorization of the manufacturer.
Type of protection (electrical):
•
MDRC-2324 (option SNIB, STIB, HNIB and HTIB):
Monitor with internal power supply: Class I equipment.
•
MDRC-2324 (option HNEW, HTEW and HPEW):
Monitor with external power supply: Class I equipment.
Degree of safety (flammable anesthetic mixture):
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or
nitrous oxide.
The equipment shall not be operable when the air oxygen content is above 25%.
Non-patient care equipment
•
Equipment primarily for use in a health care facility that is intended for use where contact with a patient is
unlikely (no applied part).
•
The equipment shall not be used with life support equipment.
•
The user should not touch the equipment, nor its signal input ports (SIP)/signal output ports (SOP) and the
patient at the same time.
Mission critical applications
We strongly recommend there is a replacement monitor immediately available in mission critical applications.
Use of electrical surgical knives
Provide as much distance as possible between the electrosurgical generator and other electronic equipment
(such as monitors). An activated electrosurgical generator may cause interference with them and can disrupt
the functionality of the display.
Power connection
–
Equipment with internal power supply
•
This equipment must be earthed.
•
Power requirements: The equipment must be powered by the AC mains voltage.
•
The equipment is intended for continuous operation.
Valid for monitors: MDRC-2324 (option SNIB, STIB, HNIB and HTIB).
Summary of Contents for Eonis MDRC-2324
Page 4: ...R5910550 00 Eonis 4...
Page 8: ...R5910550 00 Eonis 8 Welcome...
Page 12: ...R5910550 00 Eonis 12 T W Lmin Lmax Image 2 6 Installation...
Page 13: ...13 R5910550 00 Eonis Operation 3...
Page 20: ...R5910550 00 Eonis 20 Operation...
Page 21: ...21 R5910550 00 Eonis Maintenance 4...