R5910550 /00
Eonis
31
Immunity test
IEC 60601 test levels
Compliance level
Electromagnetic
environment
–
guidance
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at 0°
70% residual voltage for
25 periods at 0°
Voltage interruptions: 0%
residual voltage for 250
periods at 0°
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at 0°
70% residual voltage for
25 periods at 0°
Voltage interruptions: 0%
residual voltage for 250
periods at 0°
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the MDRC-2324 requires
continued operation
during power mains
interruptions, it is
recommended that the
MDRC-2324 be powered
from an uninterruptible
power supply or a battery
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m
Not applicable
2
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment
Conducted RF
IEC 61000-4-6
3 Vrms (6 Vrms in ISM
bands)
150 kHz to 80 MHz
3 Vrms (6 Vrms in ISM
bands)
-
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
10 V/m
Immunity to RF wireless communications equipment
Test
frequency
(MHz)
Band (MHz)
Service
Modulation
Maximum
power (W)
Distance
(m)
Immunity
test level (V/
m)
385
380
–
390
TETRA 400
Pulse
modulation
18 Hz
1.8
0.3
27
450
430
–
470
GMRS 460,
FRS 460
FM ± 5 kHz
deviation
1 kHz sine
2
0.3
28
710
704
–
787
LTE Band
13, 17
Pulse
modulation
217 Hz
0.2
0.3
9
745
780
810
800
–
960
GSM 800/
900, TETRA
800, iDEN
820, CDMA
850, LTE
Band 5
Pulse
modulation
18 Hz
2
0.3
28
870
930
1720
1700
–
1990
GSM 1800,
CDMA 1900,
GSM 1900,
DECT, LTE
Band 1/3/4/
25, UMTS
Pulse
modulation
217 Hz
2
0.3
28
1845
1970
2:
MDRC-2324 doesn
’
t contain susceptible components to magnetic fields
Summary of Contents for Eonis MDRC-2324
Page 4: ...R5910550 00 Eonis 4...
Page 8: ...R5910550 00 Eonis 8 Welcome...
Page 12: ...R5910550 00 Eonis 12 T W Lmin Lmax Image 2 6 Installation...
Page 13: ...13 R5910550 00 Eonis Operation 3...
Page 20: ...R5910550 00 Eonis 20 Operation...
Page 21: ...21 R5910550 00 Eonis Maintenance 4...