R5910550 /00
Eonis
25
Power connection
–
Equipment with external 12 VDC power supply
•
Power requirements: The equipment must be powered using the delivered medical approved 12 VDC (
) power supply.
•
The medical approved DC (
) power supply must be powered by the AC mains voltage.
•
The power supply is specified as a part of the ME equipment or combination is specified as an ME system.
•
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective
earth.
•
The equipment should be installed near an easily accessible outlet.
•
The equipment is intended for continuous operation.
Valid for monitors: MDRC-2324 (option HNEW, HTEW and HPEW).
Transient over-voltage
To fully disengage the power to the device, please disconnect the power cord from the AC inlet.
Connections
Any external connection with other peripherals must follow the requirements of clause 16 of IEC60601-1 3rd.
Ed. or Table BBB.201ofIEC 60601-1-1 for the medical electrical systems.
To maintain compliance with EMC Regulation, use only shielded interface cables for the connection to
peripheral devices.
Power cords:
•
Europe: H05VV-F or H05VVH2-F PVC cord with appropriate EU plug.
US and Canada:
“
hospital grade
”
cord-set has to be used, provided with instructions to indicate that
grounding reliability can be achieved only when the equipment is connected to an equivalent receptacle
marked hospital only or hospital grade. These instructions need to be marked either on the equip. or on a
tag on the power cord
•
Do not overload wall outlets and extension cords as this may result in fire or electric shock.
•
Mains lead protection: Power cords should be routed so that they are not likely to be walked upon or
pinched by items placed upon or against them, paying particular attention to cords at plugs and
receptacles.
•
The power supply cord should be replaced by the designated operator only at all time.
•
Use a power cord that matches the voltage of the power outlet, which has been approved and complies
with the safety standard of your particular country.
Grounding reliability
Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle.
Liquids and moisture
Never expose the monitor to liquids or moisture.
Never use the monitor near water - e.g. near a bathtub, washbasin, swimming pool, kitchen sink, laundry tub
or in a wet basement.
All models are IP20 compliant.
MDRC-2324 (option HPEW): The glass cover is fully cleanable.
Moisture condensation
Do not use monitor under rapid temperature and humidity change condition or avoid cold air from air-
conditioning outlet directly.
Moisture may condense on the surface or inside of the unit, or create a mist residue inside the protection
plate, this is not a malfunction of the product itself, although it may cause damage to the monitor.
If condensation happens, let the monitor stand unplugged until there is no condensation.
Summary of Contents for Eonis MDRC-2324
Page 4: ...R5910550 00 Eonis 4...
Page 8: ...R5910550 00 Eonis 8 Welcome...
Page 12: ...R5910550 00 Eonis 12 T W Lmin Lmax Image 2 6 Installation...
Page 13: ...13 R5910550 00 Eonis Operation 3...
Page 20: ...R5910550 00 Eonis 20 Operation...
Page 21: ...21 R5910550 00 Eonis Maintenance 4...