Bante520 Portable Conductivity Meter 

 

 

     
1.4 

Place the electrode (and temperature probe) into the conductivity standard solution, the meter 

 

automatically shows current calibration standard (e.g., 1413µ S/cm). 

 

1.5 

Press the 

Enter

 key, the meter begins the calibration, the Calibration icon will continuously   

 

flashing. 

 

1.6 

Wait for the reading to stabilize, the display automatically shows END. Single point calibration 

 

is completed.   

 

Multi-point calibration 

2.1    Ensure that you have selected 2 or 3 points calibration in the setup menu. 
2.2 

Repeat steps 1.2 to 1.5 

above, when the first calibration point is completed, the display will show “CAL2”. The meter prompts you to continue with 

second point calibration. 

2.3 

Rinse and place the electrode (and temperature probe) into the standard solution, the display automatically shows current calibration standard 

 

(e.g.,12.88mS/cm). 

2.4 

Press the 

Enter

 key, the meter begins the calibration, the Calibration icon will continuously flashing.   

2.5 

Wait for the reading to stabilize, the display will show “CAL3”. The meter prompts you to continue with third point calibrati

on. 

2.6 

Repeat the steps 2.3 and 2.4 above until the meter returns to the measurement mode. Calibration is completed. 

 

Exit the calibration 

During the calibration process, if you want to exit the calibration, press the 

Meas

 key. 

 
 

Summary of Contents for Bante520

Page 1: ...Bante520 Portable Conductivity Meter Instruction Manual BANTE INSTRUMENTS CO LTD...

Page 2: ...vironment meets following conditions Relative humidity is less than 80 Ambient temperature is greater than 0 C and less than 60 C No potential electromagnetic interference The following list describes...

Page 3: ...ent mode icon Indicates the meter is in the measurement mode Low battery alarm When the battery is depleted the icon will disappear Calibration mode icon Indicates the meter is in the calibration mode...

Page 4: ...conds Stores current reading to memory Increase value or scroll up through the menu item Views the calibration report or data logs Decrease value or scroll down through the menu item Confirms the cali...

Page 5: ...cable for power supply NOTE take out the batteries Connecting the Sensors Take out the conductivity electrode from the packaging Insert the 6 pin connector into the connector socket on meter Ensure t...

Page 6: ...for ultra pure water measurements Enable Disable Disable Set the normalization temperature for measurement and calibration 25 C 25 C 20 C Set the default temperature unit Degrees Celsius C Degrees Fah...

Page 7: ...s to enter the setup menu 2 Press the or key to select the menu item e g CAL P 02 3 Press the Enter key the meter shows an option 4 Press the or key to select the desired option press the Enter key to...

Page 8: ...ic temperature compensation mode Manual Temperature Compensation If the meter does not detect a temperature probe the C icon will show on the display indicating that the meter is switched to the manua...

Page 9: ...to the sample value you are measuring The meter will automatically detect these conductivity standard solutions and prompt the user to calibrate the meter When the calibration is completed all new ca...

Page 10: ...e when the first calibration point is completed the display will show CAL2 The meter prompts you to continue with second point calibration 2 3 Rinse and place the electrode and temperature probe into...

Page 11: ...ELE P 2 3 2 Press the Enter key The meter shows the calibration point 1 e g 10 00 S cm 3 3 Press the key the meter shows the calibration factor of the calibration point 1 e g 0 991 3 4 Press the key...

Page 12: ...uring the setting process press the or key once the setting value will increase or decrease by 0 1 Press and hold the or key the setting value will increase or decrease by 1 Conductivity Measurement 1...

Page 13: ...ading into the memory the Memory icon appears on the display Viewing stored readings 1 Press the MR key in the measurement mode the meter shows LOC P 1 Data Log 2 Press the Enter key the meter shows t...

Page 14: ...ectrode does not suit the current sample Replace the conductivity electrode Electrode is broken Replace the conductivity electrode Keypad is not working properly Replace the batteries Specifications C...

Page 15: ...in 1 litre deionised water Stir the solution until the reagent has thoroughly mixed Preparation is completed Addendum 2 How to calculate the temperature coefficient To determine the temperature coeff...

Page 16: ...to speed up these efforts Warranty The warranty period for meter is one year from the date of shipment Above warranty does not cover the sensor and calibration solutions Out of warranty products will...

Reviews: