Bante520 Portable Conductivity Meter 

14 

 

 

Connector 

6-pin 

Display 

LCD 

Power Requirements 

3 ×  

1.5V “AA” batteries

 

Dimensions 

170 (L) ×  85 (W) ×  30 (H)mm 

Weight 

300g 

 

Addendum 1: Preparation of Conductivity Standard Solutions 

Place AR potassium chloride reagent in a 50ml beaker and dry in an oven for 3 to 5 hours at 105° C, then cool to room temperature in desiccator. 

 

CALIBRATION SOLUTIONS 

REAGENT 

84µ S/cm 

Accurately weigh out 42.35mg of KCL and dissolve in 1 litre deionised water. 

1413µ S/cm 

Accurately weigh out 745.9mg of KCL and dissolve in 1 litre deionised water. 

12.88mS/cm 

Accurately weigh out 7.45g of KCL and dissolve in 1 litre deionised water. 

111.8mS/cm 

Accurately weigh out 74.5g of KCL and dissolve in 1 litre deionised water. 

 

Stir the solution until the reagent has thoroughly mixed. Preparation is completed. 
 

Addendum 2: How to calculate the temperature coefficient 

To determine the temperature coefficient of sample solution use the formula below: 

 
 
T

c

 =                                                                × 100% 

 
 
Where: 
T

= Temperature coefficient 

C

TA 

= Conductivity at Temperature A 

C

TB 

= Conductivity at Temperature B 

T

= Temperature A 

T

= Temperature B 

 
1. 

Press and hold the 

o

C

 key for 3 seconds to enter the temperature setting. 

2. 

Press the 

 or 

 key to set the temperature to 25

 

° C. 

3. 

Place the electrode into the sample solution A and record the temperature value T

A

 and conductivity value C

TA

4. 

Condition the sample solution and electrode to a temperature that is about 5

 

° C to 10

 

° C different from T

A

. Record the temperature value T

B

 and 

conductivity value C

TB

5. 

Calculate the temperature coefficient according to the formula above. 

 
 
 
 
 

C

TB

-C

TA

 

C

TA

(T

B

-25)-C

TB

(T

A

-25)

 

Summary of Contents for Bante520

Page 1: ...Bante520 Portable Conductivity Meter Instruction Manual BANTE INSTRUMENTS CO LTD...

Page 2: ...vironment meets following conditions Relative humidity is less than 80 Ambient temperature is greater than 0 C and less than 60 C No potential electromagnetic interference The following list describes...

Page 3: ...ent mode icon Indicates the meter is in the measurement mode Low battery alarm When the battery is depleted the icon will disappear Calibration mode icon Indicates the meter is in the calibration mode...

Page 4: ...conds Stores current reading to memory Increase value or scroll up through the menu item Views the calibration report or data logs Decrease value or scroll down through the menu item Confirms the cali...

Page 5: ...cable for power supply NOTE take out the batteries Connecting the Sensors Take out the conductivity electrode from the packaging Insert the 6 pin connector into the connector socket on meter Ensure t...

Page 6: ...for ultra pure water measurements Enable Disable Disable Set the normalization temperature for measurement and calibration 25 C 25 C 20 C Set the default temperature unit Degrees Celsius C Degrees Fah...

Page 7: ...s to enter the setup menu 2 Press the or key to select the menu item e g CAL P 02 3 Press the Enter key the meter shows an option 4 Press the or key to select the desired option press the Enter key to...

Page 8: ...ic temperature compensation mode Manual Temperature Compensation If the meter does not detect a temperature probe the C icon will show on the display indicating that the meter is switched to the manua...

Page 9: ...to the sample value you are measuring The meter will automatically detect these conductivity standard solutions and prompt the user to calibrate the meter When the calibration is completed all new ca...

Page 10: ...e when the first calibration point is completed the display will show CAL2 The meter prompts you to continue with second point calibration 2 3 Rinse and place the electrode and temperature probe into...

Page 11: ...ELE P 2 3 2 Press the Enter key The meter shows the calibration point 1 e g 10 00 S cm 3 3 Press the key the meter shows the calibration factor of the calibration point 1 e g 0 991 3 4 Press the key...

Page 12: ...uring the setting process press the or key once the setting value will increase or decrease by 0 1 Press and hold the or key the setting value will increase or decrease by 1 Conductivity Measurement 1...

Page 13: ...ading into the memory the Memory icon appears on the display Viewing stored readings 1 Press the MR key in the measurement mode the meter shows LOC P 1 Data Log 2 Press the Enter key the meter shows t...

Page 14: ...ectrode does not suit the current sample Replace the conductivity electrode Electrode is broken Replace the conductivity electrode Keypad is not working properly Replace the batteries Specifications C...

Page 15: ...in 1 litre deionised water Stir the solution until the reagent has thoroughly mixed Preparation is completed Addendum 2 How to calculate the temperature coefficient To determine the temperature coeff...

Page 16: ...to speed up these efforts Warranty The warranty period for meter is one year from the date of shipment Above warranty does not cover the sensor and calibration solutions Out of warranty products will...

Reviews: