Temperature Calibration
The A series meter is supplied with a TP-10K temperature probe for
measurement and temperature compensation. If the measured
temperature reading differs from that of an accurate thermometer,
the
probe needs to be calibrated.
The dissolved oxygen electrode installed with a built-in temperature
sensor and do not need to use this probe.
Connecting the Temperature Probe
1.1 Place the temperature probe into the circular hole located at the
center of the electrode arm.
1.2 Insert the jack plug to the connector socket labeled ° C. Ensure
the connector is fully seated.
Calibrating the Temperature Probe
2.1 Place the temperature probe in a solution with a known accurate
temperature and wait for the reading to stabilize.
2.2 Press and hold the
° C
key to enter the temperature setting.
2.3 Press the
/
key to modify the temperature value, press the
Enter
key to save.
To exit the calibration without saving changes, press the
ESC
key.