A150/151
I
30
1.3 Press the
Cal
key, the meter shows "Calibration Point 1" and
waits for recognizing the standard solution.
1.4 Place the electrode (and temperature probe) into the standard
solution, stir gently to remove air bubbles trapped in the slot of
the sensor.
The meter will automatically show current calibration standard
(e.g., 1413 µ S/cm). If necessary, press the
/
key to modify
the calibration value.
1.5 Press the
Enter
key, the status bar shows "
Calibrating…
"
1.6 Wait for the conductivity reading to stabilize, the meter will show
"Calibration is completed" and return to the measurement mode.
Multipoint Calibration
2.1 Ensure that you have selected 2 to 3 points calibration in the
setup menu. When the first calibration point is completed, the
screen will show "Calibration Point 2", the meter prompts you
to continue with second point calibration.
2.2 Rinse the conductivity electrode with distilled water, then rinse
with a small amount of standard solution.
2.3 Repeat the steps 1.4 and 1.5 above until the meter returns to the
measurement mode. Calibration is completed.
•
Performing the conductivity calibration will simultaneously
calibrate the corresponding TDS, salinity, resistivity and
conductivity ash values.
•
To exit the calibration without saving changes, press the
ESC
key.
Viewing the Calibration Log
3.1 Press the
MR
key, the screen shows a Data Log Menu.
3.2 Press the
Enter
key to view the calibration report.
3.3 Press the
ESC
key to return to the measurement mode.
If the meter is not calibrated, the calibration report will be unavailable.