Bante904P Portable Conductivity/DO Meter
6
Switching the Meter On and Off
•
Press the
Meas
key to switch on the meter, the display shows the measured value.
•
Press and hold the
Meas
key for 5 seconds, the meter will switch off.
Setup Menu
The Bante904P portable conductivity/dissolved oxygen meter contains an integrated setup menu that is used to customize the displayed option to meet
measurement requirements. In the different modes, the meter will show the corresponding options. The following table describes the functions of the
menu items.
CONDUCTIVITY/TDS/SALINITY/RESISTIVITY MODE:
MENU
DESCRIPTION
OPTIONS
DESCRIPTION
DEFAULT
Cell Constant:
Set the cell constant to match connected electrode.
K=0.1
K=1
K=1
K=10
User-Defined
Temperature Coefficient:
Set the linear temperature compensation coefficient.
Range: 0.0~10.0%/° C
2.10
Calibration Points:
Set the number of calibration points.
1 point
1 point
2 points
3 points
4 points
5 points
Pure Water Coefficient:
When the option is enabled, pure water coefficient
will be applied automatically for ultra-pure water
measurements.
Enable
Disable
Disable
Reference Temperature:
Set the normalization temperature for conductivity
measurement and calibration.
25° C
25° C
20° C
TDS Factor:
Set the default TDS conversion factor.
Range: 0.40~1.00
0.5
Measurement Unit:
Set the default temperature unit.
Degrees Celsius
° C
Degrees Fahrenheit