Bante904P Portable Conductivity/DO Meter
12
Conductivity Calibration - Manual Calibration
The Bante904P portable conductivity/dissolved oxygen meter provides an easy manual calibration mode. If the conductivity standard solution is not ready,
you are able to use this method for calibration.
3.1
Record the cell constant value on the conductivity electrode (e.g., K=1.01).
3.2
Press and hold the
key for 3 seconds to enter the setup menu, the display shows
“CELL
/P-01
”.
3.3
Press the
Enter
key, the meter enters the setting mode.
3.4
Press the
or
key
until the display shows “User/Cell”, press the
Enter
key to confirm.
3.5
Press the
or
key to set each digit according to previous record.
3.6
Press the
Enter
key to confirm until the setting values stop flashing. The meter returns to the measurement mode. Calibration is completed.
•
Performing the conductivity calibration will simultaneously calibrate the corresponding TDS, salinity and resistivity values.
•
If you want to exit the calibration, press the
Meas
key.