Bante904P Portable Conductivity/DO Meter
11
1.4
Place the electrode into the conductivity standard solution, the meter automatically shows
current calibration standard (e.g., 1413µ S/cm).
1.5
If necessary, press the
or
key to set the calibration value.
1.6
Press the
Enter
key, the Calibration icon begins flashing.
1.7
Wait for the reading to stabilize, the display automatically shows END. Single point calibration
is completed.
Multi-point calibration
2.1 Ensure that you have selected 2 or 5 points calibration in the setup menu.
2.2
Repeat steps 1.2 to 1.6
above, when the first calibration point is completed, the display will show “CAL2”. The meter prompts you to continue with
second point calibration.
2.3
Rinse and place the electrode into the standard solution, the display automatically shows current calibration standard (e.g.,12.88mS/cm).
2.4
If necessary, press the
or
key to set the calibration value.
2.5
Press the
Enter
key, the Calibration icon begins flashing.
2.6
Wait for the reading to stabilize, the display will show “CAL3”. The met
er prompts you to continue with third point calibration.
2.7
Repeat the steps 2.3 and 2.5 above until the meter returns to the measurement mode. Calibration is completed.