3.2
Ultrasound signals
- not applicable -
3.3
Special functions
Degas (
– in the time scale area)
• The DEGAS function is switched on for degassing the system before sonication. To do so,
set the required time with the preset time button, then hold the Start/ Stop button for at least
2 seconds.
The system can be stopped prematurely by pressing the Start/Stop button.
During the degassing, the top green LED (
) also
fl
ashes until the time expires.
• Switching between ultrasound – degas: If the Start/ Stop button is kept depressed for a long
time when ultrasound is running, the ultrasound is
fi
rstly switched off and then after approx. 2
seconds the Degas function is reactivated.
Locking continuous operation ( – in the time scale area)
• To avoid accidentally turning on the continuous operation, the continuous operation can be
deactivated:
Pull the mains plug out.
Press the preset time button, hold it depressed and insert the mains plug at the same time.
The yellow LED "1 min" lights up in con
fi
rmation.
The function can be reactivated in the same way. The green LED continuous operation ( ) lights
up in con
fi
rmation.
4
Use
Depending on the application, the samples are sonicated
indirectly
in laboratory
fl
asks or other sample vessels, which are inserted in the
contact
fl
uid in the oscillating tank.
Direct
sonication of objects in the bath
fl
uid in the ultrasonic baths DT
... F is not permitted.
The intended course is described step by step in the following sections.