background image

14 / 32

51326c GB/2017-06

1

 electronic GmbH & Co. KG  •  Heinrichstraße 3-4  •  12207 Berlin  •  Deutschland  •  [email protected]

4.1  

Fill  oscillating tank  

If inset beakers are used in the  ultrasonic bath, a contact 

fl

 uid must always be used as otherwise 

the ultrasound cannot pass through the air (between the ultrasonic oscillating tank and the inset 
beaker). This  contact 

fl

 uid must be able to transfer the ultrasound to the  

fl

 uid in the inset beaker 

without loss if possible. The surface tension is very high in municipal and deionised water. As a 
result of this high surface tension, the cavitation does not occur throughout the entire solution. 
It generally only forms on the bottom of the oscillating tank or on inhomogeneities in the water. 
To reduce the surface tension, you should add a surfactant preparation which is biodegradable, 
easily disposable, gentle to the material and long-lasting. We recommend  TICKOPUR TR 3, which 
is easily available from specialist laboratory supply dealers. Even when used for a long period 
of time, this preparation does not leave any scrum or incrustations,does not turn cloudy and on 
average functions for 5-7 days before needing to be changed. 

The ultrasonic bath is turned off for  

fi

 lling. 

  Check that the  ball valve is closed. 

  Fill 1/3 of oscillating tank with water. 

  Add the surfactant preparation in the correct dosage, e.g., 1% 

 TICKOPUR TR 3. 

 Dosage

Filling 

quantity

 TICKOPUR TR 3

1%

Water

+

Concentrate

DT 510 F

2.5 litres

2.48 l

+

25 ml

DT 1028 F

5.8 litres

5.74 l

+

58 ml

  Fill carefully up to the  

fi

 lling level mark, avoid as much as possible the 

formation of foam. The oscillating tank must be 

fi

 lled depending on the 

sample vessels used, as they displace the  contact 

fl

 uid.

IMPORTANT:

With the sample vessels inserted, the 

fi

 lling level should be at or slightly above the  

fi

 lling level 

mark, as a too low 

fi

 lling level damages the ultrasonic bath! 

Instructions –  

fi

 lling

• Do not 

fi

 ll oscillating tank with hot water. Maximum  

fi

 lling temperature: 50°C.

•  At least drinking-quality water must be used to 

fi

 ll the oscillating tank.

•  Only use distilled or deionised water without additives in inset beakers or insert tubs. 
•  Do not use any combustible, explosive, non-aqueous  liquids or azeotropic mixtures directly in 

the stainless steel oscillating tank (e.g. benzine, solvents). Furthermore, chemicals that contain 
or that separate chloride ions (some disinfectants, household cleaners, and dish detergents), 
may not be used directly in the stainless steel tank.

•  When working with aggressive preparations in inset beakers or insert tubs: Prevent the  contact 

fl

 uid or stainless steel surfaces from being sprayed. If necessary, replace the  contact 

fl

 uid, clean 

the surfaces and wipe dry.

Summary of Contents for Sonorex Digitec DT 510 F

Page 1: ...3 D Series F Ultrasonic baths for preparation of samples User instructions Valid for DT 510 F and DT 1028 F 51326c GB 2017 06...

Page 2: ...g and has no legal effect In case of any discrepancies between the translation and the original of this document the original version has priority BANDELIN accepts no responsibility or liability for d...

Page 3: ...expire If any service is required please contact the specialist dealer in charge or the manufacturer Symbols used Symbol Significance Explanation Danger Identifies information that could signify a ri...

Page 4: ...1 Scope of delivery 11 2 2 Set up assembly 11 2 3 Start up 11 3 Operation 12 3 1 Operating elements 12 3 1 1 Ultrasound 12 3 2 Ultrasound signals not applicable 13 3 3 Special functions 13 4 Use 13 4...

Page 5: ...ecks 19 6 3 Error analysis 19 6 4 Repairs and service 20 6 4 1 Decontamination certificate 20 6 4 2 Exchanging fuses 20 7 Accessories 21 7 1 Required accessories 21 7 2 Optional accessories not applic...

Page 6: ...they contain piezoelectric transducers the energy of which is transferred to the bath fluid with ultrasound frequency as mechanical oscillations As a result microscopically small bubbles are continuou...

Page 7: ...supplied beaker holder or another holder is required for storing the inset beakers with the samples during the sonication Nothing must stand or lie directly on the bottom of the bath this is the only...

Page 8: ...II Degree of contamination 2 Permissible ambient temperature 5 to 40 C Permissible relative humidity up to 31 C 80 Permissible relative humidity up to 40 C 50 No dewing Only for indoor operation Spec...

Page 9: ...280 320 10 1 4 F2A In order to improve the effect the ultrasound is being modulated whereby depending upon the ultrasonic bath a 4 or 8 fold Ultrasonic nominal output value is yielded as ultrasonic pe...

Page 10: ...nded socket Do not operate the ultrasonic bath without fluids Do not stand or lay any objects on the tank bottom accessories must be used see Section 7 Do not immerse any parts of the body e g hands f...

Page 11: ...rder see delivery note 2 2 Set up assembly Place the ultrasonic bath atop a firm level and dry surface In doing so observe the maximum weight of the ultrasonic bath including fluid Net weight see Tech...

Page 12: ...the time scale Following pressing of the Start Stop button a running light displays the remaining time optically Once the time expires the ultrasonic output is stopped automatically Premature pressin...

Page 13: ...essed for a long time when ultrasound is running the ultrasound is firstly switched off and then after approx 2 seconds the Degas function is reactivated Locking continuous operation in the time scale...

Page 14: ...e is closed Fill 1 3 of oscillating tank with water Add the surfactant preparation in the correct dosage e g 1 TICKOPUR TR 3 Dosage Filling quantity TICKOPUR TR 3 1 Water Concentrate DT 510 F 2 5 litr...

Page 15: ...is recommended When using preparations the safety instructions included in the product leaflets are to be fundamentally adhered to Replace used sonication fluids do not refresh by adding fluids 4 2 De...

Page 16: ...low the vessels The fill level of the oscillating tank must always be at or slightly above the filling level mark following equipment A low fill level will damage the ultrasonic bath Set the required...

Page 17: ...o be extended with acidic cleaning solutions The cavitation noise changes during degassing loud degassing noises disappear at the end of the degassing process the ultrasonic bath appears to work more...

Page 18: ...te the passive protective layer of the stainless steel activating the stainless steel and making it begin to rust The extraneous rust produces localised corrosion of the stainless steel For this reaso...

Page 19: ...value in the technical data Section 1 4 tolerances 20 Checking the ultrasound effect For monitoring we recommend the performance of a foil test biannually A suitable frame for the foil test can be re...

Page 20: ...tion is to be filled out and affixed to the packaging on the outside in a visible spot If this form has not been filled out we reserve the right to refuse receipt of the package in order to protect ou...

Page 21: ...the tank bottom 7 2 Optional accessories not applicable 7 3 Preparations Special preparations are required for use of the ultrasound that are ultrasound compatible i e they are cavitation conducive b...

Page 22: ...10 11 15 16 20 21 Ambient conditions 8 B Ball valve 6 11 14 17 Bath fluid 6 7 10 13 15 16 17 18 Beaker holder 7 11 13 16 21 C Care 18 Care instructions 17 CE conformity 3 7 Cleaning 18 Cleaning proce...

Page 23: ...7 8 Mode of operation 6 O Operating elements 10 12 Operating location 11 Operating temperature 16 Outlet 6 9 P Preparation of samples 7 Preset time 12 13 Protection class 8 Purpose 7 8 R Regulations 1...

Page 24: ...Ultrasound Operation 10 16 18 Ultrasound output 12 Use 13 V VAH certified 18 W Warning instructions 10 BANDELIN electronic GmbH Co KG Heinrichstra e 3 4 12207 Berlin Deutschland www bandelin com info...

Page 25: ......

Page 26: ...pwards To be screwed into insert baskets and utensil holders EK 10 10 ml max 31 mm EK 25 25 ml max 42 mm EK 50 50 ml max 52 mm EK 100 100 ml max 65 mm EK 250 250 ml max 85 mm Beaker holder GL F stainl...

Page 27: ...Accessories Devices Holding clamps for laboratory flasks Beaker holder Shaking device DT 510 F EK 10 EK 25 EK 50 EK 100 EK 250 GL 510 F DT 1028 F EK 10 EK 25 EK 50 EK 100 EK 250 2 GL 510 F SA 1028...

Page 28: ...e foils to be compared at any time The perfo rated areas of all foils should have approx the same dimensions and distribution the results are never identical A process validation e g for the treatment...

Page 29: ...contact the manu facturer Type for interior tank dimensions in mm L x W x H Order No for from to FT 1 190 85 60 3190 DT 31 H DT 52 H RK 31 H RK 52 H FT 4 240 140 80 300 150 150 3074 DL 102 H DL 255 H...

Page 30: ...current laws and regulations and if necessary must also be disinfected with a surface disinfection agent listed by the VAH Device type Serial No Accessories Device accessories are not contaminated wer...

Page 31: ...ns and that the information provided in this declaration are correct and complete Company Institute Street and number Postal code city Department Name Telephone extension Fax Reason for return Thank y...

Page 32: ...Note The user instructions in this and other languages as well as further information are available on the supplied CD...

Reviews: