Dismounting
Mounting
Sickle spanner
HS 40/42
Spanner
here: Spanner SW 10
29 / 42
51376-012 en/2022-09
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
4.3.1.2 Mounting and dismounting of the titanium probes
Titanium probes (TS ...) are mounted directly on the UW 100/200 on the standard/booster
horn SH 100 G/SH 200 G or on the UW 50.
First, carefully wipe the mounting surfaces of the standard/booster horn and
the probe
ensure clean mounting surfaces.
Screw on the probe by hand.
Lay the probe on an approx. 20 mm thick base so that it does not bend.
Firmly hold the standard/booster horn in place using sickle spanner HS 40/42, and tighten
the probe using the appropriate spanner, see illustration. The use of the torque wrench
DMS 10 is recommended.
Note: Observe the permissible torque!- see product information
With the UW 50 use the sickle spanner HS 25/28 - the probe is mounted directly
onto the UW 50.
Dismount in the reverse order.