Error
Possible cause
Remedial action
Signi
fi
cant heating in the
vicinity of the mounting
surfaces between the
ultrasonic converter and
standard/booster horn or
standard/booster horn –
probe?
Sonicating parts (standard/booster
horn and probe) not mounted
fi
rmly enough?
Are mounting surfaces of the
sonicating parts soiled?
- Dismount the respective parts, clean
the surfaces and
fi
rmly retighten, see
chapter 4.4.
If it is not possible to rectify the error using these short instructions, please contact your local
specialist dealer or write to us at the address below.
6.4
Repairs and service
If you identify errors or defects that cannot be recti
fi
ed, the ultrasonic homogeniser may no longer
be used.
If this occurs, please contact the supplier or the manufacturer, use the enclosed questionnaire.
BANDELIN electronic GmbH & Co. KG
Heinrichstrasse
3-4
12207
Berlin
Repairs/ Maintenance Department:
E-mail:
Tel.:
+49-(0)-30 - 768 80 - 13
Fax:
+49-(0)-30 - 76 88 02 00 13
Returns are subject to the General Conditions of Delivery and Payment of
BANDELIN electronic GmbH & Co. KG.
In addition, the ultrasonic homogeniser must be cleaned and decontaminated (if necessary),
see the following chapter.
Caution
•
Only allow repairs to be conducted by authorised skilled personnel!
• Before each repair, turn o
ff
the device and disconnect the plug from the mains!
•
Defective parts may only be replaced with original parts.
• The black housing of the ultrasonic converter and the aluminium cylinder (ultrasonic oscillating
system) may not be twisted toward each other. The ultrasonic oscillating system and its
electrical connections would be damaged as a result.