Step 3 :
Switch on the ultrasonic homogeniser
Switch on the ultrasonic homogeniser at the mains switch.
All saved data and settings will be loaded. Check the settings and load a di
ff
erent program
if necessary, see chapter 3.3.
Step 4 :
Activate the ultrasound
Start with a small amplitude to avoid spraying. Observe the maximum allowed amplitude,
see chapter 4.3.
Activate the ultrasound.
Warning: risk of splashing!
This is especially the case with small sample quantities and when immersing oscillating probes.
Step 5 :
Remove the sample
After sonication, the probe must be removed from the medium. Leaving it in the sonicating
medium for a longer time can cause damage to the probe.
Once the program or time setting has elapsed, the ultrasound emission ends automatically.
If continuous sonication has been set, the ultrasound emission must be stopped manually.
Remove the probe and the temperature sensor, if present, from the sonicating medium.
Step 6 :
Clean the probe
In order to prevent contamination with other sonicating media, probes are to be thoroughly
cleaned after every sonication.
Switch
o
ff
the ultrasonic homogeniser at the mains switch.
Clean the probe and check the wear on the sound emitting surface at regular intervals,
see chapter 5.1.