Dismounting
Ultrasonic transducer
Standard horn
Vigorously turn sickle spanner
HS 40/42 long, in the opposite
direction
30 / 48
51376g GB/2018-04
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
4.4
Dismounting and mounting the standard horns
Standard horns
• are screwed onto the respective ultrasonic transducer,
• are made of high-strength titanium alloy,
• transmit the oscillations to the probe,
• enhance the amplitude thanks to their geometry.
Standard horn SH 100 G is
fi
rmly mounted onto the UW 100 and SH 200 G onto UW 200 at the
time of delivery.
Before mounting the standard horns, the HF generator must be turned off and the ultrasonic
transducer must be disconnected from the HF generator.
Release
the
fi
rmly-mounted standard horn SH 100 G from the ultrasonic transducer UW 100.
To do so, lay the ultrasonic transducer on a
fi
rm base (non-slip if possible).
Dismounting
Place one sickle spanner on the ultrasonic transducer and the other sickle spanner on the
standard horn. To dismount, press the
fi
rst sickle spanner against the
fi
rm base, press the
other sickle spanner
fi
rmly in the opposite direction.
One
sickle spanner is part of the s cope of delivery of one standard horn.