Dismounting
Assembly
Sickle spanner
HS 40/42
Spanner
here: Spanner SW 10
29 / 48
51376g GB/2018-04
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
4.3.1.2 Mounting and dismounting of titanium probes
• Titanium probes (TS ...)
-
On UW 100/200, titanium probes are mounted onto standard horn SH 100 G/SH 200 G.
On UW 50 they are directly mounted.
First, carefully wipe the mounting surfaces of the standard horn and the probe
ensure
clean mounting surfaces.
Screw on the probe by hand.
Lay the probe on an approx. 20 mm thick base so that it does not bend.
Firmly hold the standard horn in place using sickle spanner HS 40/42, and tighten the probe
using the appropriate spanner, see illustration. It is recommended to the use the torque
wrench DMS 10.
Note: Observe the torque.
With the UW 50 use sickle spanner HS 25/28 - the probe is mounted directly onto the UW 50.
Dismount in the reverse order.