Action
Display
Remark
8
Cancelling
Press the RESET key
MIN/SEC:
last set value
CYCLE:
newly set value
POWER: 000
%
The current program can be
cancelled at any time by pressing
the RESET key. The last set program
will be automatically displayed again.
If the RESET key is pressed once
again after the cancellation, the
generator will switch to continuous
operation.
9
Switching to continuous
operation when the
generator has stopped
Press the RESET key 1 time
MIN/SEC:
00 min 00 sec
CYCLE:
newly set value
POWER: 000
%
10
Calling up a saved program
when the program has
expired
Press the RESET key once.
MIN/SEC: saved
value
CYCLE: saved
value
POWER: 000
%
Press once:
the saved program will appear
in continuous operation
Press twice:
the saved program will appear
in countdown operation
11
Turning off the generator
Turn the control dial all the
way to the left. Activate the
mains switch.
The display goes out.
The control light is no longer lit.
The last set program is retained.
After powering off, a pause of
approx. 10 sec must be observed
before turning the device on again.
The saved countdown operation is
retained.
Caution ! Observe the maximum allowed amplitude setting for the probes - chapter 4.3.
19 / 44
51436a GB/2017-10
1