Assembly
Ultrasonic transducer
Standard and/or booster horn
Vigorously turn sickle spanner
HS 40/42 long, or HS 45/50 long,
in the opposite direction
Plastic ruler
Threaded bolt
Assembly
Keep the surfaces/threads to be screwed clean; clean them with alcohol and a lint-free cloth, if
needed.
Screw approx. 2/3 of the length of the threaded bolts of the new standard or booster horn to
be mounted, into the mounting surface of the ultrasonic transducer.
Place a plastic ruler or similar on the threaded portion of the bolt and press lightly in the
direction of the arrow
⇒
. This will prevent the bolt from turning any further when the standard
or booster horn is screwed on.
Screw the standard or booster horn onto the threaded bolt and
fi
rst tighten by hand after removing
the ruler.
Then, mount the standard or booster horn
fi
rmly onto the ultrasonic transducer using both
sickle spanners.
Detailed information on the individual standard and/or booster horns can be found in the separate
product information.
The black housing of the ultrasonic transducer and the aluminium cylinder (ultrasonic oscillating
system) must not be twisted toward each other. The ultrasonic oscillating system and its electrical
connections would be damaged as a result.