• Only use the prescribed tool for mounting and dismounting (see appendix B).
• Do not use any bent probes (
⇒
unstable operation, loss of power).
•
Liquids must not penetrate the inside of the ultrasonic transducer.
• Never twist the ultrasonic transducer's black housing toward the aluminium cylinder (ultrasonic
oscillating system). The ultrasonic oscillating system and its electrical connections would be
destroyed as a result.
• Do not operate the device without supervision.
Advice for the medical and laboratory
fi
eld
• The
homogeniser is exclusively intended for use by skilled medical personnel.
• It may generate radio interferences or disrupt the operation of devices nearby. It may be
necessary to take remedial measures such as realigning the device or recon
fi
guring the
homogeniser or the shield.
• During
operation, portable or mobile HF communication systems in the vicinity of the
homogeniser should be turned off - operation may be disrupted.
Damage
• If damage to the homogeniser is detected, do not connect the homogeniser to the mains.
• In the event of defects, disconnect the power plug immediately.
•
Repairs are only to be conducted by authorised skilled personnel or by the manufacturer.
•
Defective parts must only be replaced with original parts or parts of the same quality!
2 Preparation
Carefully unpack the HF generator, ultrasonic transducer and accessories, and inspect them for
completeness or possible transportation damage. If any damage or defects are found, these must
be immediately noti
fi
ed in writing to the transportation company and to the supplier.
Before operation, the ultrasonic bath should be allowed to stand for 2 hours at its operating
location so that it can adjust to the ambient conditions.