4.3
Selecting a suitable probe
Detailed information on the individual probes can be found in the separate product information. A
graphic overview of accessory parts can be found in appendix A.
HD 2070
Probe
Ø
MS 72
2 mm
MS 73
3 mm
KE 76
6 mm
TT 13
13 mm
VS 70 T
13 mm
TT 13 FZ
13 mm
Connection
to standard
horn
SH 70 G, (FZ 5 G)
FZ 5 G
Sonicating
vol.
1 ml -
25 ml
2 ml -
50 ml
5 ml -
100 ml
10 ml -
200 ml
10 ml -
200 ml
-
Amplitude
285
μ
m
ss
245
μ
m
ss
191
μ
m
ss
93
μ
m
ss
97
μ
m
ss
93
μ
m
ss
Maximum
permitted
amplitude
setting
97 %
100 %
100 %
100 %
100 %
100 %
HD 2200
Probe
Ø
MS 72
2 mm
MS 73
3 mm
KE 76
6 mm
TT 13
13 mm
VS 70 T
13 mm
TT 13 FZ
13 mm
TT 19
19 mm
VS 190 T
19 mm
TT 25
25 mm
VS 200 T
25 mm
Connection
to standard
horn
SH 213, SH 213 G, (FZ 7 G)
FZ 7 G
SH 219 G
SH 225 G
Sonicating
vol.
2 ml -
30 ml
5 ml -
90 ml
10 ml -
350 ml
20 ml -
900 ml
20 ml -
900 ml
-
25 ml -
900 ml
25 ml -
900 ml
30 ml -
1000 ml
30 ml -
1000 ml
Amplitude
286
μ
m
ss
308
μ
m
ss
255
μ
m
ss
165
μ
m
ss
170
μ
m
ss
165
μ
m
ss
81
μ
m
ss
81
μ
m
ss
53
μ
m
ss
51
μ
m
ss
Maximum
permitted
amplitude
setting
52 %
65 %
72 %
100 %
100 %
100 %
100 %
100 %
100 %
100 %
21 / 44
51436a GB/2017-10
1