Cut the hose to a suitable length and connect it between the laboratory cooler and
the 3-way ball valve.
The hose must not be kinked.
Fix the outlet at the tank edge of the ultrasonic bath with the screw. Connect the hose
to the laboratory cooler.
Mounting – ultrasonic bath inlet
Place the inflow nozzle on a tank corner of the ultrasonic bath and fasten it with
the screws
Cut the hose to a suitable length and connect it between the laboratory cooler and
the inlet nozzle.
The hose must not be kinked.
2.3
Putting into service
In order to transmit the ultrasound between the oscillating tank and the laboratory
fl
asks,
the tank must be
fi
lled with a contact liquid composed of normal municipal or deionised
water together with a surfactant, e.g. TICKOPUR TR 3.
Connect the laboratory cooler to the power supply (grounded socket).
Fill the oscillating tank of the ultrasonic bath according to the separate operating
instructions. Set/open the 3-way ball valve on the ultrasonic bath in the direction of
the drain hose.
The liquid must not be contaminated with particles.
Oscillating
tank
fi
lled to
fi
lling level mark.
Switch on the laboratory cooler at the main switch – the pump starts immediately.
It may be necessary to
fi
ll the oscillating tank up to the
fi
lling mark.