443 M1 01 E
10-03-2015
TECHNICAL FEATURES
Output signal
4-20 mA ; 4 wire technique ; compatible with BAMOPHOX 436 (230 V AC version)
Power supply:
24 V DC (10 ... 30 V DC), galvanic insulated
Consumption:
about 0.8 W; if the 4-20 mA loop is separately powered the consumption is 0.5 W
CAUTION:
if a connection is done with another instrument than BAMOPHOX 436 (230 V AC version)
a separated power supply could be necessary according to power supply available on the PLC.
Ambient temperature: + 0°C … 45°C
Immersion:
10 m as a maximum
Sensor protection:
IP68 acc. EN 60 529 (max. 10 m) Sensor PVC immersion probe ; connection to head housing: IP67 connector
Media temperature:
+0°C … 60°C
Ranges:
1 = 0...50 FNU / 2 = 0...100 FNU / 3 = 0...200 FNU / 4 = 0...500 FNU / 5 = 0...1000 FNU / 6 = 0...2000 FNU
Accuracy:
±5 % of reading and ±1 % full scale of actual range in use
Resolution:
0.01 – 1 FNU (depending on range in use)
Housing:
PBT Head, protection IP 65 (EN 60 529)
Fitting connection BSP 2” / bracket and blocking nut PG9
Display:
All inside the head housing: 4 digits display for reading measured turbidity, programming and settings
LED status on alarms and functions
Configuration:
Through rotating selector and push button
Cable:
Length 6 m (other on request)
CE label:
According to Directives
2006/95/CE
and
2004/108/EG
FNU = Formazin Nephelometric Units
Ø 93
215
40
40
63
70
Pg 9
M20x1.5
156
48
310