443 M1 01 E

10-03-2015

TECHNICAL FEATURES 

Output signal  

4-20 mA ; 4 wire technique ; compatible with BAMOPHOX 436 (230 V AC version)  

Power supply:   

24 V DC (10 ... 30 V DC), galvanic insulated  

Consumption:   

about 0.8 W; if the 4-20 mA loop is separately powered the consumption is 0.5 W  

CAUTION: 

if a connection is done with another instrument than BAMOPHOX 436 (230 V AC version)
a separated power supply could be necessary according to power supply available on the PLC. 

Ambient temperature:   + 0°C … 45°C  
Immersion:   

10 m as a maximum  

Sensor protection:   

IP68 acc. EN 60 529 (max. 10 m) Sensor PVC immersion probe ; connection to head housing: IP67 connector 

Media temperature:  

+0°C … 60°C

Ranges:  

1 = 0...50 FNU  /  2 = 0...100 FNU  /  3 = 0...200 FNU  /  4 = 0...500 FNU  /  5 = 0...1000 FNU  /  6 = 0...2000 FNU

Accuracy:   

±5 % of reading and ±1 % full scale of actual range in use  

Resolution:   

0.01 – 1 FNU (depending on range in use)  

Housing:    

PBT Head, protection IP 65 (EN 60 529)     
Fitting connection BSP 2” / bracket and blocking nut PG9

Display:    

All inside the head housing: 4 digits display for reading measured turbidity, programming and settings     
LED status on alarms and functions  

Configuration:   

Through rotating selector and push button  

Cable:    

Length 6 m (other on request)    

CE label:

According to Directives 

2006/95/CE

and 

2004/108/EG

FNU = Formazin Nephelometric Units

Ø 93

215

40

40

63

70

Pg 9

M20x1.5

156

48

310