FUNCTIONS – DESCRIPTION

- TURBISENS determines the turbidity of liquids according ISO 7027, using infrared light in pulsating light mode with 2 emitters and

1 receiver.

- Output signal: 4 to 20 mA
- Automatic compensation is assured on external light and colour.

MOUNTING : Caution!

The probe must be installed vertically using the supplied bracket in order to let a 

free space of 10 cm

as a minimum 

from the tank or channel floor. The cable sensor is connected with an IP67 protected connector to the head housing. 

RANGE SET UP 

Connect the power supply, at this moment the initializing 

sequence runs until the green LED blinks.

Press the push button. Display: 

"OUT" 

Press the push button once more. Display: 

"present value" 

Choose the right range with the selector. 

The instrument after 5 seconds saves your selected range 
and returns back to measuring mode. 

See fig.1

: Configuration menu

Switching the power on:

The  initializing  sequence  begins,  display  indicates  the  sofware
version.
Whitin 3 seconds, green LED blinks: measuring mode is opera-
ting and display indicates the measured turbidity. 

Compensation:

Colour compensation only for 0...50 and 0...100 FNU ranges. 

Status indicators: 

Blinking green LED: 

= All in operation, measurement mode 

OFF green LED: 

= Faulty power supply 

ON red LED: 

= Dysfunction on 4-20 mA output 

Error indications on the display: 

"SUN"

–> Too much external light 

"dlrt"

–> Dirty glass windows

"           "

–> Over range

"noSd"

–> Faulty probe or not connected,

broken cable

Dysfunction diagnostics 

- No entry on colour compensation:

This function is available only for two ranges 0...50 and 
0...100 FNU. 
Choose an appropriate range to check this function. 

- No choice of range over 100 FNU: 

The colour compensation is working and does not allow any
range over 0...100 FNU. 
Set off the colour compensation and set up the right range. 

Monitor

+ -

+

www.ier.de

Enter

- +

-

4-20mA

10-30V

Power

Output

DC

LED status

Display

parameters

and turbidity

Rotating selector for

parameters selection

and 

Push button for

parameters entry

Galvanic

insulated

output

Input

Power supply

Fig.1

SAFETY PRECAUTIONS

• Only qualified person should operate, install and assure the maintenance. All International and European recommendations on electrical 

installation should be applied.

• The device may only be connected to supply power, which complies with the specifications included in the technical data.
• The device must be disconnected from all sources of power during installation and maintenance work. 
• The device may only be operated under the conditions specified in the operating instructions. 