TESTING THE TURBISENS WITH CALIBRATION KIT
Factory Calibration:
The sensor is factory calibrated. As a rule, no recalibration is required.
Test Equipment Monitoring:
If device calibration testing is required as part of the respective quality assurance system for
test equipment monitoring, calibration can be checked with the calibration kit. This kit is pai-
red with only one TURBISENS: refer to serial numbers.
Test procedure
●
Lift up the TURBISENS out of liquid.
●
Clean, rinse and dry carefully the beam windows (even traces of drop must be removed).
Do not use abrasive materials.
●
Set up the measuring range to 200.00 FNU.
●
Fit the test kit on the probe as shown on the picture.
There is no specific way to slide it on the probe.
●
After 15 s the instrument is ready for check-up.
●
Compare the value in FNU engraved on the kit with the reading value
- If reading is 15% under the kit value
–> Calibration test is right.
- If reading is 15% over the kit value
–> send the instrument to BAMO
for a factory calibration
TURBISENS Test-Unit
SN:
1234567
FNU:
56
Date:
01.02.09
CAUTION
TURBISENS and calibration kit
Must have the same serial number
Calibration Kit identification
TURBISENS