TESTING THE TURBISENS WITH CALIBRATION KIT

Factory Calibration:

The sensor is factory calibrated. As a rule, no recalibration is required.

Test Equipment Monitoring:

If device calibration testing is required as part of the respective quality assurance system  for
test equipment monitoring, calibration can be checked with the calibration kit.  This kit is pai-
red with only one TURBISENS: refer to serial numbers. 

Test procedure 

Lift up the TURBISENS out of liquid.  

Clean, rinse and dry carefully the beam windows (even traces of drop must be removed).
Do not use abrasive materials. 

Set up the measuring range to 200.00 FNU.  

Fit the test kit on the probe as shown on the picture.  
There is no specific way to slide it on the probe.  

After 15 s the instrument is ready for check-up. 

Compare the value in FNU engraved on the kit with the reading value 

- If reading is 15% under the kit value  

–> Calibration test is right. 

- If reading is 15% over the kit value

–> send the instrument to BAMO  

for a factory calibration 

TURBISENS Test-Unit

SN:

1234567

FNU:

56

Date:

01.02.09

CAUTION

TURBISENS and calibration kit 

Must have the same serial number 

Calibration Kit identification 

TURBISENS 