194 M1 01 F

12-13-2013

1.  TECHNICAL FEATURES

Displayed parameters: 

Measured values - Configuration Menu - Temperature 

Display: 

Back lighted - 2 lines of 16 alphanumerical characters ; 9 mm high

Indication: 

LED alarms status

Configuration: 

8 push buttons keyboard on front board - Keyword protected

Sensor input:

For amperometric sensors 4-20 mA 

(2 wire technique)

or 0/-2 V 

(4 wire technique)

Scales: 

Configuration within the sensor in use, from 0.001 to 9999 ppm or g/L

Accuracy: 

depending of the sensor in use 

(see technical features of sensor)

Flow control

Input for inductive sensor type NPN 

(making contact if present flow rate)

Temperature:

±3.0 °C

Relay outputs: 

4 contacts (Silver alloy), voltage free

Thresholds: 

3 programmable independent thresholds - with adjustable hysteresis 0…100 %

threshold S3 available in copy of external input signal 

(example: flow sensor)

and adjustable, timer from 0 to 9999 sec

Output relay (S4) 

Common alarm signal for: Too long injection, Temperature out of range etc. 

Contact: 

Initial resistance 0.1 Ohm as a maximum (voltage drop 6 V DC 1 A)

Rated at 831 V AC / 3 A / 277 V AC ; 90 W / 3 A / 30 V DC

Switching capacity 

(minimum)

100 mA, 5 V DC

(depending of switching frequency, ambient conditions, accuracy)

Mechanical life time 

(minimum)

5 x10

6

operations 

(180 commutation /min)

Electrical life time 

(minimum)

2 x10

5

(20 comm./min)

[3 A, 125 V AC], [3 A, 30 V DC] 

and 10

5

(evaluated charge) for 3 A, 125 V AC

ON/OFF Regulation: 

Pulse time 0…9999 sec - High and low proportional bandwidth, high and low dead zone.

Calibration sequence: 

Regulation on standby, relay outputs inhibited, analogical outputs stand on last values

Measurement output: 

0/4-20 mA (maxi 600 Ohm) proportional to measurement, galvanic insulated

Temperature output: 

0/4-20 mA (max 600 Ohm), scaling -20 to +160°C, galvanic insulated

Program Testing: 

simulation through the menu on measurement, temperature, and relay outputs

Main power supply: 

230 V AC / 50-60 Hz (other on request) - Consumption 10 VA

Models: 

Panel mounting, IP65, 72 x 144 mm, connections on screw terminal IP40

Idem DIN Rail mounting, only for blind monitor

Wall mounting, IP65, cable glands, connections on screw terminal

OPTION (RS 422 + Logger)

Communication: 

RS422 output, J-BUS link, binary slave mode, 2400 to 9600 bauds

Data Logger: 

Cycle average measurement record, with a programmable period,

150 000 records on memory card / External driver necessary for reading

2.  DIMENSIONS 

Extension terminal:

identical to the panel
or wall mounting

PANEL

Wall mounting instrument

Panel mounting instrument

Summary of Contents for BAMOPHOX 194

Page 1: ...MES 194 01 1 Amperometric signal monitor BAMOPHOX 194 13 12 2013 194 M1 01 F INSTRUCTIONS MANUAL BAMOPHOX 194 Amperometric sensors signal monitor...

Page 2: ...ONT PANEL 7 SCROLLING MENU 8 ABOUT BAMOPHOX 9 CONSULTING MODIFYING 9 TESTING RELAY OUTPUTS 9 SENSOR CONFIGURATION 10 ADJUSTING THRESHOLD 1 11 ADJUSTING THRESHOLD 2 11 ADJUSTING THRESHOLD 3 12 ON OFF R...

Page 3: ...C depending of switching frequency ambient conditions accuracy Mechanical life time minimum 5 x106 operations 180 commutation min Electrical life time minimum 2 x105 20 comm min 3 A 125 V AC 3 A 30 V...

Page 4: ...M1 01 F 12 13 2013 WALL MOUNTINg INSTRUMENT 3 WIRINg OPTION LOGGER RS 422 OPTION LOGGER access by upper cover RS 422 PANEL MOUNTINg INSTRUMENT 29 30 31 36 V V 0V E Signal input Signal input 29 30 31 3...

Page 5: ...ntact Measurement Temperature S3 threshold NO contact Measurement Temperature External sensor NOT USED Ground N Main power supply N Neutral L Line L V V 0V E 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17...

Page 6: ...SENSOR CONNECTION Load resistor supplied with the BAMOPHOX 194 4 20 mA sensor 0 0 to 2000 mV sensor TERMINAL FOR SENSOR CONNECTION Flow rate sensor 194 832 NPN type normally open Brown Blue Black outp...

Page 7: ...his key all parameters are saved previous data programmation will be overwrited If you are not sure of any modification do not press the VALID key To change parameters of data capture Num ric input in...

Page 8: ...ADJUST ALARM 1 ADJUST ALARM 2 ADJUST ALARM 3 RELAY REGULATION REGUL PID OUTPUT mA MES OUTPUT mA TEMP TEMPERATURE SENSOR CALIBRATION FORCED RELAY ADJUST ALARM CLOCK RECORDING TIME LANGUAGE MENU MENU M...

Page 9: ...the configuration thresholds regulation analog outputs When PID regulation is activated the display shows the PID FORCED PID 00 00 ppm 00 00 One digit is flashing Modify the value Immediately the inst...

Page 10: ...D regulation with the 4 20 mA signal There is a specificity of alarm 3 in Bamophox 194 It can be setup for an external sensor Signal is coming from an inductive flow rate sensor type NPN This sensor...

Page 11: ...e contact S1 SAVINg ADJUST ALARM 2 ALARM 2 ON OFF ALARM 2 MEASURE TEMP MEASURE Threshold against measured value TEMP Threshold against temperature measured value ALARM 2 LOW HIgH HIgH Contact closes w...

Page 12: ...he limit LOW Contact closes when the value goes under the limit ON 00 00 ppm To close the contact S3 at this value OFF 00 00 ppm To open the contact S3 at this value DELAY UP ONI OFF Delay or no delay...

Page 13: ...est value for LOW PB 00 00 ppm proportional bandwith Highest value H IGH DZ 00 00 ppm for dead zone Lowest value LOW DZ 00 00 ppm for dead zone SAVING ENTER ENTER ENTER ENTER ENTER ENTER ENTER ENTER E...

Page 14: ...tion contact S1 is closed Inside the dead zone between 1 4 1 9 ppm S1 and S2 are OFF no injection at all contacts S1 and S2 are opened If the measurement value is between the dead zone and the highest...

Page 15: ...such as m Cde t delay of response T Time corresponding to the same variation in of measurement m and the analog output Cde m Cde This value may be found out on the slope Modify the PID parameters as f...

Page 16: ...Page 16 194 M1 01 F 12 13 2013 ppm measurement...

Page 17: ...lue corresponding to 00 00 or 04 00 mA OUTPUT 4 20 mA 0 20 mA Output type SAVING ENTER ENTER ENTER ENTER VALID TEMPERATURE AUTO A Pt 100 probe should be connected MEASURE AUTO MANUAL MANUAL Without Pt...

Page 18: ...ARD is displayed Proceed at the same time to a test to know the standard value according to the sensor in use Enter the result value For good calibration value has to be as closed as the sensor full r...

Page 19: ...g a malfunction on S1 and S2 contacts an overtime contact could be set up WITH WHITHOUT ALARM TMAX S1 0000 Sec Over timing on S1 closed contact maximum time for active relay TMAX S2 0000 Sec Over timi...

Reviews: