Page 17
194 M1 01 F
12-13-2013
Output mA MEASUREMENT
Measurement signal copy on the analog output
HIGHER
00.00
ppm
Value corresponding to 20,00 mA
LOWER
00.00
ppm
Value corresponding to 0.00 or 04,00 mA
OUTPUT
4/20 mA
/
0/20 mA
Output type
SAVING ?
▲
▲
▲
▲
▲
ENTER
ENTER
ENTER
ENTER
VALID
Output mA TEMP.
HIGHER
0000
°C
Value corresponding to 20.00 mA
LOWER
0000
°C
Value corresponding to 00.00 or 04.00 mA
OUTPUT
4/20 mA
/
0/20 mA
Output type
SAVING ?
▲
▲
▲
▲
▲
ENTER
ENTER
ENTER
ENTER
VALID
TEMPERATURE
AUTO
: A Pt 100 probe should be connected
MEASURE :
AUTO / MANUAL
MANUAL
: Without Pt 100 probe
Manualy entered temperature value
FLUIDE T. +
0000
°C
(In MANUAL Mode only)
SAVING ?
▲
▲
▲
ENTER
ENTER
ENTER
VALID
Caution: When PID regulation is active, this step menu does not appears
MENU
+
Next menu:
Output mA TEMP.
MENU
+
Next menu:
TEMPERATURE
MENU
+
Next menu:
SENSOR CALIBRATION
To switch the PID regulation on stand-by,
input 24 V= 20 mA on terminals 5(+) and 6(-).