background image

Page 19

194 M1 01 F

12-13-2013

FORCED RELAY

ALARM 1

ON / OFF  

ALARM 2

ON / OFF   

Diagnostic mode to test manually the threshold configurations

ALARM 3

ON / OFF   

ALARM 4

ON / OFF 

ENTER

ENTER

ENTER

ENTER

ENTER

ADJUST ALARM

ENABLES or DISABLES the S4 relay

(In case of malfunction of the control relay)

This mode allows detecting a malfunction on S1
and S2 contacts ; an overtime contact could be set up.

WITH / WHITHOUT 

ALARM

TMAX. S1  

0000

Sec

Over timing on S1 closed contact

(maximum time for active relay)

TMAX. S2  

0000

Sec

Over timing on S2 closed contact

(maximum time for active relay)

SAVING ?

ENTER

ENTER

ENTER

ENTER

VALID

LANgUAgE

FRENCH / ENGLISH / ITALIAN
GERMAN

SAVING ?

ENTER

ENTER

VALID

NOTE:

To complete the configuration

of a 

BAMOPHOX Logger

(option) 

follow the corresponding manual.

MENU

+

Next menu:

ADJUST ALARM

MENU

+

Next menu:

LANGUAGE             or   CLOCK

Summary of Contents for BAMOPHOX 194

Page 1: ...MES 194 01 1 Amperometric signal monitor BAMOPHOX 194 13 12 2013 194 M1 01 F INSTRUCTIONS MANUAL BAMOPHOX 194 Amperometric sensors signal monitor...

Page 2: ...ONT PANEL 7 SCROLLING MENU 8 ABOUT BAMOPHOX 9 CONSULTING MODIFYING 9 TESTING RELAY OUTPUTS 9 SENSOR CONFIGURATION 10 ADJUSTING THRESHOLD 1 11 ADJUSTING THRESHOLD 2 11 ADJUSTING THRESHOLD 3 12 ON OFF R...

Page 3: ...C depending of switching frequency ambient conditions accuracy Mechanical life time minimum 5 x106 operations 180 commutation min Electrical life time minimum 2 x105 20 comm min 3 A 125 V AC 3 A 30 V...

Page 4: ...M1 01 F 12 13 2013 WALL MOUNTINg INSTRUMENT 3 WIRINg OPTION LOGGER RS 422 OPTION LOGGER access by upper cover RS 422 PANEL MOUNTINg INSTRUMENT 29 30 31 36 V V 0V E Signal input Signal input 29 30 31 3...

Page 5: ...ntact Measurement Temperature S3 threshold NO contact Measurement Temperature External sensor NOT USED Ground N Main power supply N Neutral L Line L V V 0V E 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17...

Page 6: ...SENSOR CONNECTION Load resistor supplied with the BAMOPHOX 194 4 20 mA sensor 0 0 to 2000 mV sensor TERMINAL FOR SENSOR CONNECTION Flow rate sensor 194 832 NPN type normally open Brown Blue Black outp...

Page 7: ...his key all parameters are saved previous data programmation will be overwrited If you are not sure of any modification do not press the VALID key To change parameters of data capture Num ric input in...

Page 8: ...ADJUST ALARM 1 ADJUST ALARM 2 ADJUST ALARM 3 RELAY REGULATION REGUL PID OUTPUT mA MES OUTPUT mA TEMP TEMPERATURE SENSOR CALIBRATION FORCED RELAY ADJUST ALARM CLOCK RECORDING TIME LANGUAGE MENU MENU M...

Page 9: ...the configuration thresholds regulation analog outputs When PID regulation is activated the display shows the PID FORCED PID 00 00 ppm 00 00 One digit is flashing Modify the value Immediately the inst...

Page 10: ...D regulation with the 4 20 mA signal There is a specificity of alarm 3 in Bamophox 194 It can be setup for an external sensor Signal is coming from an inductive flow rate sensor type NPN This sensor...

Page 11: ...e contact S1 SAVINg ADJUST ALARM 2 ALARM 2 ON OFF ALARM 2 MEASURE TEMP MEASURE Threshold against measured value TEMP Threshold against temperature measured value ALARM 2 LOW HIgH HIgH Contact closes w...

Page 12: ...he limit LOW Contact closes when the value goes under the limit ON 00 00 ppm To close the contact S3 at this value OFF 00 00 ppm To open the contact S3 at this value DELAY UP ONI OFF Delay or no delay...

Page 13: ...est value for LOW PB 00 00 ppm proportional bandwith Highest value H IGH DZ 00 00 ppm for dead zone Lowest value LOW DZ 00 00 ppm for dead zone SAVING ENTER ENTER ENTER ENTER ENTER ENTER ENTER ENTER E...

Page 14: ...tion contact S1 is closed Inside the dead zone between 1 4 1 9 ppm S1 and S2 are OFF no injection at all contacts S1 and S2 are opened If the measurement value is between the dead zone and the highest...

Page 15: ...such as m Cde t delay of response T Time corresponding to the same variation in of measurement m and the analog output Cde m Cde This value may be found out on the slope Modify the PID parameters as f...

Page 16: ...Page 16 194 M1 01 F 12 13 2013 ppm measurement...

Page 17: ...lue corresponding to 00 00 or 04 00 mA OUTPUT 4 20 mA 0 20 mA Output type SAVING ENTER ENTER ENTER ENTER VALID TEMPERATURE AUTO A Pt 100 probe should be connected MEASURE AUTO MANUAL MANUAL Without Pt...

Page 18: ...ARD is displayed Proceed at the same time to a test to know the standard value according to the sensor in use Enter the result value For good calibration value has to be as closed as the sensor full r...

Page 19: ...g a malfunction on S1 and S2 contacts an overtime contact could be set up WITH WHITHOUT ALARM TMAX S1 0000 Sec Over timing on S1 closed contact maximum time for active relay TMAX S2 0000 Sec Over timi...

Reviews: