Page 12
194 M1 01 F
12-13-2013
ADJUST ALARM 3
ALARM 3
ON/OFF
EXTERN
NO/YES
NO
= set up (ALARM 3 MEASURE/TEMP)
YES
= flow rate sensor wired
SAVINg ?
ALARM 3
MEASURE/TEMP
MEASURE
= Threshold against measured value
TEMP.
= Threshold against temperature measured value
ALARM 3 LOW/HIgH
HIgH
= Contact closes when value goes over the limit
LOW
= Contact closes when the value goes under the limit
ON
00.00
ppm
To close the contact S3 at this value
OFF
00.00
ppm
To open the contact S3 at this value
DELAY UP
ONI/OFF
Delay (or no delay) before to close the contact S3
TIME
0000
SEC
Delay time to close the contact S3
DELAYDOWN
ON/OFF
Delay (or no delay) before to open the contact S3
TIME
0000
SEC
Delay time to open the contact S3
SAVINg ?
MENU
+
Next menu:
RELAY REGULATION
▲
▲
▲
▲
▲
▲
▲
▲
▲
▲
▲
▲
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
VALID
VALID