Page 56 of 59
Excessive Gas Usage
If you have a warning on the touchscreen for excessive gas usage as shown in Screen 19
we recommend you carry out the following to try to rectify the issue:
Set the workstation to the main menu, Screen 3
Remove the glove port caps and wipe around the rim of the caps and the internal
seal with an appropriate cleaning agent
Check that the glove port tube connector is fitted into the floor of the workstation
at the front right
Ensure that the inner interlock door is firmly shut and that the door clamps are
holding it in please
If the workstation is fitted with the ultrasonic Humidity option check that the main
water tank is fitted and that the 2 tubes are fitted correctly
Check that the gas bottle pressures are set appropriately as detailed in the Gas
Supply Requirements section
Additionally you can also remove the front screen to wipe around the internal edge of the
screen and the seal of the workstation.
If the problem persists after the above have been carried out then please contact your
distributor to assist in identifying the issue or to arrange a service engineer visit.
Internal Equipment not powering up
If 3
rd
party equipment being used in the workstation are not powering up then check the
following:
Check the power / current rating of the equipment. There is a 2A limit for internal
equipment and if the equipment exceeds this then it will likely blow an internal
fuse. If this is the case then you will need to contact your local distributor for repair
Check that the internal sockets are switched on through either Screen 10 or Screen
15
Check that both ends of the mains cable are fully inserted into their sockets
If the above do not work then try to power the equipment from a wall socket to check that
it is working properly. If it does work correctly outside of the workstation then contact your
local distributor or Ruskinn for further assistance.
The Workstation is only partially operational
If some features of the workstation are not working correctly then please contact your
local distributor to arrange a service engineer visit.